background image

Chapter 8 - Replacement Procedures

 8-95

 
D

IRECTION

 5771498-100, R

EVISION

 6

V

ENUE

™ S

ERVICE

 M

ANUAL

P R E L I M I N A R Y

8-6-6

Casters Replacement Procedure

NOTE:

The procedure below is the same whether replacing a No Lock Caster or a Directional and Brake Lock 
Caster

NOTE:

Make sure to install each type in its correct location.See 

Figure 8-82

.

8-6-6-1

Tools

Allen key 6mm

LOCTITE® 243™

8-6-6-2

Time Required

15 min

8-6-6-3

Preparations

Shut down the Venue™  ultrasound unit, as described in 

Power Shut Down 

.

Make sure the system is standing securely on a level surface, with the wheels in the locked position.

8-6-6-4

Casters Removal Procedure

1) Using a 6mm Allen key, loosen and remove the screw that secures the wheel shaft in the wheel 

securing socket.

2) Lift the chassis sufficiently to allow the wheel to drop down out of the wheel securing socket.

3.) Remove the wheel.

8-6-6-5

Casters Installation Procedure

1) Carefully lift the chassis sufficiently to allow insertion of the replacement wheel shaft into the wheel 

securing socket.

Note:

When installing the swivel lock wheel, lock the swivel before installing it to determine the 
installation direction.

2) Push the wheel shaft all the way up into the socket, then gently lower the chassis to the ground.

3.) Apply LOCTITE® 243™ on the Allen screw.

FRU Part #

Refer to 

Table 9-7

 on page 9-6

.

Figure 8-81   Removing the Caster Wheels

Release the screw

Caster Wheels Disconnected

Summary of Contents for Venue R1

Page 1: ...Operating Documentation 2019 General Electric Company 16 JUN 2019 GEHC_FRNT_CVR FM P R E L I M I N A R Y Venue Service Manual Direction Number 5771498 100 Revision 6 Technical Publication...

Page 2: ...2019 General Electric Company Manufacturer GE Medical Systems Ultrasound Primary Care Diagnostic LLC 9900 Innovation Drive Wauwatosa WI 53226 USA 16 JUN 2019 GEHC_FRNT_CVR FM P R E L I M I N A R Y...

Page 3: ...tandard Hazard Icons 1 7 Safety Considerations 1 8 Introduction 1 8 Human Safety 1 8 Mechanical Safety 1 11 Electrical Safety 1 13 Venue Battery Safety 1 17 Patient Data Safety 1 17 Dangerous Procedur...

Page 4: ...ments 2 6 Facility Needs 2 7 Purchaser Responsibilities 2 7 Required Facility Needs 2 8 Networking Pre Installation Requirements 2 12 Connectivity Installation Worksheet 2 13 Overview 3 1 Purpose of C...

Page 5: ...Software Options Configuration 3 42 Connectivity Overview 3 43 Physical Connection 3 43 Stand alone Venue 3 43 Sneaker Net Environment 3 43 Wireless Connection from Venue to DICOM Server 3 43 Connecti...

Page 6: ...g and Transporting the Venue Ultrasound Scanner 4 11 Functional Checks 4 14 Overview 4 14 Performance Checks 4 14 2D Mode B Mode Checks 4 14 M Mode Checks 4 15 PW CW Doppler Mode Checks 4 17 Probe Con...

Page 7: ...19 Signal Flow 5 19 Front End Power Supply T FEPS 5 20 Front End T CFE Board 5 20 Probe Selection Board T PSB 5 21 Back End Processor 5 22 Introduction 5 22 Back End Processing BEP Module 5 22 BEP Bl...

Page 8: ...wivel Resistance Adjustments 6 4 Arm Vertical Movement Adjustments 6 4 Arm Rotate Resistance Adjustments 6 5 Overview 7 1 Purpose of Chapter 7 7 1 Service Safety Considerations 7 2 Service Tools 7 3 V...

Page 9: ...oting 7 25 Purpose of this Section 7 25 MPB Diagnostics Procedure Troubleshooting 7 25 Overview 8 1 Purpose of Chapter 8 8 1 Visual Guide 8 1 Accessories Replacement Procedures 8 2 Basic Storage Baske...

Page 10: ...ocedure 8 71 T TRx Box Replacement Procedure 8 73 T TRx Module Replacement Procedure 8 75 Mechanical Parts Replacement Procedures 8 78 MPB Front Metal Door Replacement Procedure 8 78 MPB Guide L and G...

Page 11: ...Procedures 8 184 Burning Disk on Key Media with SW Downloaded from GE Portal 8 184 Software Update Procedure 8 188 Software Upgrade Procedure from Venue R1 to Venue R2 8 193 Software Installation Pro...

Page 12: ...Schedule 10 3 Tools Required 10 5 Tools Required for Servicing the Venue 10 5 System Maintenance 10 6 Preliminary Checks 10 6 Functional Checks 10 7 Physical Inspection 10 9 Cleaning 10 10 Probe Main...

Page 13: ...DIRECTION5771498 100 REVISION 6 VENUE SERVICE MANUAL Table of Contents 13 P R E L I M I N A R Y This page was intentionally left blank...

Page 14: ...S UNE AUTRE LANGUE QUE L ANGLAIS C EST AU CLIENT QU IL INCOMBE DE LE FAIRE TRADUIRE NE PAS TENTER D INTERVEN TION SUR LES QUIPEMENTS TANT QUE LE MANUEL SERVICE N A PAS T CONSULT ET COMPRIS LE NON RESP...

Page 15: ...A DO T CNICO OPERADOR OU PACIENTE DEVIDO A CHOQUES EL TRICOS MEC NICOS OU OUTROS ESTE MANUAL DE ASSIST NCIA EST DISPON VEL APENAS EM INGL S SE QUALQUER OUTRO SERVI O DE ASSIST NCIA T CNICA QUE N O A G...

Page 16: ...UUTA KUIN ENGLANNINKIELIST MATERIAALIA TARVITTAVAN K NN KSEN HANKKIMINEN ON ASIAKKAAN VASTUULLA L YRIT KORJATA LAITTEISTOA ENNEN KUIN OLET VARMASTI LUKENUT JA YMM RT NYT T M N HUOLTO OHJEEN MIK LI T T...

Page 17: ...A POSKYTOVATELE SLU EB OBSLU N HO PERSON LU NEBO PACIENT VLIVEM ELEKTRICK HOP PROUDU RESPEKTIVE VLIVEM K RIZIKU MECHANICK HO PO KOZEN NEBO JIN MU RIZIKU DENNE SERVICEMANUAL FINDES KUN P ENGELSK HVIS E...

Page 18: ...K JAS OPERATORIUS AR PACIENTAS GALI B TI SU EISTAS D L ELEKTROS SM GIO MECHANINI AR KIT PAVOJ DENNE SERVICEH NDBOKEN FINNES BARE P ENGELSK HVIS KUNDENS SERVICELEVERAND R TRENGER ET ANNET SPR K ER DET...

Page 19: ...ORAREA ACESTUI AVERTISMENT AR PUTEA DUCE LA R NIREA DEPANATORULUI OPERATORULUI SAU PACIENTULUI N URMA PERICOLELOR DE ELECTROCUTARE MECANICE SAU DE ALT NATUR OVAJ PRIRU NIK ZA SERVISIRANJE DOSTUPAN JE...

Page 20: ...TERA ILI PACIJENTA PROUZRO ENE STRUJNIM UDAROM MEHANI KIM I DRUGIM OPASNOSTIMA T TO SERVISN PR RU KA JE K DISPOZ CII LEN V ANGLI TINE AK Z KAZN KOV POSKYTOVATE SLU IEB VY ADUJE IN JAZYK AKO ANGLI TINU...

Page 21: ...M TER TEKN SYEN KILAVUZUN NG L ZCE DI INDAK B R D LDE OLMASINI STERSE KILAVUZU TERC ME ETT RMEK M TER N N SORUMLULU UNDADIR SERV S KILAVUZUNU OKUYUP ANLAMADAN EK PMANLARA M DAHALE ETMEY N Z BU UYARINI...

Page 22: ...P R E L I M I N A R Y DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL ix 6 16 19 POC_PRECAUTIONS_LEGAL ZH CN KO...

Page 23: ...rical work on these products will comply with the requirements of the applicable electrical codes The purchaser of GE equipment shall only utilize qualified personnel i e GE s field engineers personne...

Page 24: ...change without notice Every precaution has been taken in the preparation of this document Nevertheless GE assumes no responsibility for errors omissions or any damages including special or consequent...

Page 25: ...s and corrections 3 11 FEBURARY 2018 Support R2 configuration 4 16 JULY 2018 Added Kisyu codes ECG support and Upgrade from R1 to R2 procedure 5 10 APRIL 2019 Added R2 5 DOK 6 6 JUNE 2019 Added troubl...

Page 26: ...N A R Y Chapter 1 Introduction Section 1 1 Overview 1 1 1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing the Venue ultrasound scanner The service provider mus...

Page 27: ...installation planners some parts of Chapter 2 Site Preparations Table 1 1 Contents in this Service Manual CHP NUMBER TITLE DESCRIPTION Chapter 1 Introduction Contains a content summary and warnings C...

Page 28: ...in Table 1 2 below NOTE When not specified otherwise the contents in this manual apply to all Venue models Table 1 2 Venue Models PSI Group Cat No Description PSI Code GP Code Kisyu Code Venue R1 H45...

Page 29: ...tion in 2D Color Doppler M Mode Color M Mode and PW The fully digital architecture of the Venue system allows optimal usage of all scanning modes and probe types throughout the full spectrum of operat...

Page 30: ...hardware versions on the boards Refer to HW Version tab To check the software versions on local software on the boards Refer to Software Version To check for installed options Refer to Options Setup...

Page 31: ...message Known or potential hazards to personal are labeled in one of three ways DANGER WARNING CAUTION When a hazard is present that can cause property damage but has absolutely no personal injury ris...

Page 32: ...d to make you aware of specific types of hazards that could possibly cause harm Refer to the User Manual for a complete list of icons used on the Ultrasound System not all icons may be listed in the t...

Page 33: ...OVED FROM AN OPERATING ULTRASOUND SYSTEM SOME METAL SURFACES MAY BE WARM ENOUGH TO POSE A POTENTIAL HEAT HAZARD IF TOUCHED EVEN WHILE IN SHUTDOWN MODE WARNING WARNING BECAUSE OF THE LIMITED ACCESS TO...

Page 34: ...R AT LEAST 30 SECONDS FOR CAPACITORS TO DISCHARGE AS THERE ARE NO TEST POINTS TO VERIFY ISOLATION THE LIGHT ON THE OP PANEL ON OFF BUTTON WILL TURN OFF ULTRASOUND SYSTEM COMPONENTS MAY BE ENERGIZED AL...

Page 35: ...erview P R E L I M I N A R Y CAUTION USE PROTECTIVE GLASSES DURING DRILLING FILING AND DURING ALL OTHER WORK WHERE EYES NEED PROTECTION CAUTION USE SAFETY SHOES WHEN DOING WORK WHERE THERE IS ANY CHAN...

Page 36: ...WEIGHT OF INSTALLED PERIPHERALS WHEN READY FOR USE CARE MUST BE USED WHEN MOVING IT OR REPLACING ITS PARTS FAILURE TO FOLLOW THE PRECAUTIONS LISTED BELOW COULD RESULT IN INJURY UNCONTROLLED MOTION AND...

Page 37: ...le inside the vehicle Secure the system with straps or as directed otherwise to prevent motion during transport Prevent vibration damage by driving cautiously Avoid unpaved roads excessive speeds and...

Page 38: ...connector must meet international electrical standards WARNING WARNING CONNECTING A VENUE SCANNER TO INCORRECT POWER LEVEL WILL DESTROY THE SYSTEM CONNECT THE SYSTEM ONLY IN ACCORDANCE WITH THE VOLTAG...

Page 39: ...Avoid pulling pinching or kinking the probe cable since a damaged cable may compromise the electrical safety of the probe To avoid the risk of a probe accidentally falling do not allow the probe cabl...

Page 40: ...ce a distance of 2 5 m 8 2 ft above the floor on which the medical personnel stand and a horizontal distance of 1 5 m 4 9 ft have justified themselves as indicative of the dimensions of the Patient WA...

Page 41: ...see the example in Figure 1 1 Patient Environment EN 60601 1 Sub clause 3 79 Patient Environment It is difficult for this standard to define dimensions for the volume in which diagnosis monitoring or...

Page 42: ...other heat sources Do not leave the battery in direct sunlight Do not pierce the battery with a sharp object hit it or step on it Do not use a damaged battery Do not apply solder to a battery Do not...

Page 43: ...MOVED FROM AN OPERATING VENUE SYSTEM SOME METAL SURFACES MAY BE WARM ENOUGH TO POSE A POTENTIAL HEAT HAZARD IF TOUCHED EVEN WHILE IN SHUT DOWN MODE WARNING WARNING EXPLOSION WARNING DO NOT OPERATE THE...

Page 44: ...All potentially hazardous stored or residual energy is relieved WARNING WARNING ENERGY CONTROL AND POWER LOCKOUT FOR VENUE WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER...

Page 45: ...xplanation of label icons and symbols refer to Table 1 3 on page 1 23 A system Rating Label examples shown in Figure 1 3 and Figure 1 4 below is located at the rear of the system This indicates the ul...

Page 46: ...1 Introduction 1 21 P R E L I M I N A R Y When Upgrading the software from R1 to R2 an electronic label will be displayed under the About tab in the Config screen Figure 1 5 Electronic label eLabel f...

Page 47: ...ION 5771498 100 REVISION 6 VENUE SERVICE MANUAL 1 22 Section 1 1 Overview P R E L I M I N A R Y Figure 1 7 General Label International R2 Figure 1 8 Venue Warning Label Figure 1 9 Stored Energy Hazard...

Page 48: ...user documentation when complete information cannot be provided on the label Various WARNING Dangerous voltage the lightning flash with arrowhead in equilateral triangle is used to indicate electric...

Page 49: ...t push the unit sideways when the caster wheel brakes are in the locked position Instability may occur Rating Plate DO NOT place objects on the surface of the rear of the LCD Panel while folded Rating...

Page 50: ...rded Bottom This symbol indicates the product contains hazardous materials in excess of the limits established by the Chinese standard GB T 26572 Requirements of concentration limits for certain restr...

Page 51: ...n the Assembled in X statement is not shown on the label this indicates that the Customs country of origin is the same as the country of the legal manufacturer Every system has a unique marking for id...

Page 52: ...R E L I M I N A R Y 1 7 3 Venue External Labels Location In addition to the labels described in the previous section an additional label may be found on the Venue ultrasound unit as described in the...

Page 53: ...or intended for use in patient care are regulated medical waste for transportation purposes and must be transported as a hazardous material NOTE The USER SERVICE staff should dispose of all the waste...

Page 54: ...supplied with the system NOTE Portable and mobile RF communications equipment can affect the Venue ultrasound scanner 1 9 2 Compliance The Venue ultrasound scanner conforms to all applicable conducte...

Page 55: ...IONS WARNING WARNING RISK OF ELECTRICAL SHOCK SYSTEM MUST BE TURNED OFF AVOID ALL CONTACT WITH ELECTRICAL CONTACTS CONDUCTORS AND COMPONENTS ALWAYS USE NON CONDUCTIVE HANDLES DESIGNED FOR THE REMOVAL...

Page 56: ...capture cine loop 4D multi volume loop Restart the application before resuming clinical scanning Table 1 4 Phone Numbers for Customer Assistance 1 of 2 LOCATION PHONE NUMBER USA GE Ultrasound Service...

Page 57: ...E Healthcare Pvt Ltd 4 Kadugodi Industrial Area Bangalore 560 067 India Phone 91 1 800 425 8025 91 1 800 425 7255 91 1 800 102 7750 Table 1 5 Phone and Fax Numbers for Manufacturer MANUFACTURER PHONE...

Page 58: ...r 2 This chapter provides the information required to plan and prepare for the setup of a Venue ultrasound unit Included are descriptions of the electrical and facility requirements that must be met b...

Page 59: ...needed for Ultrasound system installation updates and repairs However operator and patient comfort may be optimized if the room light is subdued and indirect Therefore a combination lighting system di...

Page 60: ...conduit 2 2 5 1 Venue Power Requirements Electrical specifications for the Venue system are as follows 2 2 5 2 Inrush Current Inrush current is not a factor for consideration due to the inrush current...

Page 61: ...here is no guarantee that interference will not occur in a particular installation NOTE Possible EMI sources should be identified before the unit is installed and should not be on the same line as the...

Page 62: ...magnet wrap at each end Install all covers Loose or missing covers or Radio Frequency gaskets allow radio frequencies to interfere with the ultrasound signals Replace broken RF gaskets If more than 20...

Page 63: ...n as possible if possible six weeks before delivery to allow enough time to make any changes WARNING WARNING FOR CONSOLE ULTRASOUND SYSTEMS AND FOR ULTRASOUND SYSTEMS MOUNTED ON A DOCKING ISOLATION CA...

Page 64: ...e electrical codes The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment The desire to use a non listed or customer provided product or t...

Page 65: ...eral are within 2 m 6 5 ft of each other with peripheral within 1 m of the Venue system to connect cables Power outlets for other medical equipment Power outlets for test equipment within 1 m 3 2 ft o...

Page 66: ...full size ground wire from the distribution panel to the Ultrasound outlet Sites with a mains power system without a defined Neutral The dedicated line shall consist of one phase two lines not shared...

Page 67: ...inet 11 Network Interface 3 Film Viewer 12 457 mm 18 inches distance of Ultrasound system from wall or objects 4 Counter Top 13 Stool 5 Counter Top and Sink with hot and cold water 14 Footswitch 6 Ove...

Page 68: ...de Circuit breaker for dedicated power outlet is easily accessible Sink with hot and cold water Receptacle for bio hazardous waste like used probe sheaths Emergency oxygen supply Storage for linens an...

Page 69: ...ervers for storage and to workstations for viewing images As an added benefit transferring images in this manner enables viewing to be done on reviewing station while scanning continues 2 3 3 4 DICOM...

Page 70: ...2 3 Connectivity Installation Worksheet Site System Information Services Destination Devices TCP IP Settings Device Type Manufacturer Name IP Address Port AE Title 1 2 3 4 5 6 7 8 9 10 11 12 Dept Venu...

Page 71: ...sheet for DICOM Network Information Venue Host Name AE Title Local Port IP Address Net Mask ROUTING INFORMATION ROUTER2 ROUTER3 ROUTER1 Destination IP Addresses GATEWAY IP Addresses Default DICOM APPL...

Page 72: ...ances from the basic setup Make sure system and probes have been subject to acclimation period Environmental cooling is sufficient Lighting is adjustable to adapt to varying operational conditions of...

Page 73: ...2 16 Section 2 4 Connectivity Installation Worksheet DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y This page was intentionally left blank...

Page 74: ...pared setup can proceed as described in this chapter Included in this chapter are guidelines for transporting the unit to a new site as well as procedures that describe how to receive and unpack the e...

Page 75: ...3 5 3 2 2 1 System Acclimation Time When unpacking the Ultrasound system allow the temperature of the Ultrasound system to stabilize before powering up The following table describes guidelines for re...

Page 76: ...THE SYSTEM AS IT IS HEAVY TWO PEOPLE ARE ALWAYS REQUIRED WHENEVER A PART WEIGHING 16KG 35 LB OR MORE MUST BE LIFTED CAUTION TO PREVENT ELECTRICAL SHOCK CONNECT THE UNIT TO A PROPERLY GROUNDED POWER O...

Page 77: ...r wooden crate mounted on a raised wooden base see Figure 3 5 on page 3 5 CAUTION Two people are needed to unpack the Ultrasound system because of its weight Attempts to move the Ultrasound system con...

Page 78: ...Venue Table 3 10 Shipping Carton Dimensions and Weights Description Height Width Depth Weighta a Weight is approximate and will vary depending upon the supplied peripherals Venue scanner with periphe...

Page 79: ...DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL 3 6 Section 3 3 Receiving and Unpacking the Equipment P R E L I M I N A R Y Figure 3 7 Shipping Packages...

Page 80: ...inspection of the shipping carton proceed to the unpacking instructions below as appropriate NOTE When using utility a knife to cut the nylon wrap take extra care not to scratch the system or any per...

Page 81: ...e Equipment P R E L I M I N A R Y 3 Remove the accessories boxes 4 Remove the protective foams from the top of the system Figure 3 9 Removing Accessories Boxes Figure 3 10 Removing Protective Foams Pr...

Page 82: ...VISION 6 VENUE SERVICE MANUAL Chapter 3 System Setup 3 9 P R E L I M I N A R Y 5 Remove the wooden ramps and then remove the nylon wrap from the system Figure 3 11 Removing Nylon Wrap from the System...

Page 83: ...iving and Unpacking the Equipment P R E L I M I N A R Y 6 Unlock the four butterfly locks and remove the two wooden brackets 7 Remove the silica gel bags Figure 3 12 Removing Wooden Brackets Figure 3...

Page 84: ...completely from the shipping carton 10 Make sure the system is standing firmly on the floor before applying the locking brake 11 Before commencing the installation procedure follow the instructions fo...

Page 85: ...rom the safety latches 2 Remove the locks and open the four safety latches located on both sides of the crate 3 The shipping crate contains the following items Venue system Power cord Manuals Accessor...

Page 86: ...DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL Chapter 3 System Setup 3 13 P R E L I M I N A R Y Figure 3 17 Shipping Crate Contents...

Page 87: ...E MANUAL 3 14 Section 3 3 Receiving and Unpacking the Equipment P R E L I M I N A R Y 4 Lay the front wooden crate door ramp and secure it with the 2 metal pins located inside the crate Figure 3 18 Wo...

Page 88: ...UE SERVICE MANUAL Chapter 3 System Setup 3 15 P R E L I M I N A R Y 6 Remove the anti static foams 7 Remove the wooden planks to gain access to the system Figure 3 19 Removing the Anti Static Foams Fi...

Page 89: ...acking the Equipment P R E L I M I N A R Y 8 Remove the plastic bags from the system 9 Release the caster locking on the front and rear casters 10 Carefully roll the scanner forwards down the unloadin...

Page 90: ...netic Interference EMI Many of the covers shields and screws are provided primarily to protect the system from image artifacts caused by this interference For this reason it is imperative that all cov...

Page 91: ...lugged Clean the console 2 Probes Check all probes for wear and tear on the lens cable and connector Look for bent or damaged pins on the connector and in the connector socket on the unit Verify that...

Page 92: ...w of the Venue Ultrasound Unit Figure 3 22 Front View of the Venue Ultrasound Scanner Item 1 Cockpit monitor Tilts up and down and swivels left and right 2 eTower 3 Basket holders 4 Caster 1x front wi...

Page 93: ...ush the scanner and place it in the desired position 10 On Off Switch Cover 3 Probe locking levers 11 Interface Panel Provides ports for LAN Insulated USB and Dual USB see Figure 3 24 4 Caster 1x back...

Page 94: ...not insulated 3 HDMI connector not insulated 3 4 4 EMI Protection The Venue has been designed to minimize the effects of Electro Magnetic Interference EMI Many of the covers shields and screws are pr...

Page 95: ...right corner at 1 30 m above the floor and in a distance of 1 meter at all four sides 1 meter above the floor 3 5 3 Electrical Specifications 3 5 3 1 Verification of the Venue Voltage Setting Verify...

Page 96: ...valid version of the system standard IEC60601 1 1 Everybody who connects additional equipment to the signal input part or signal output part of Venue configures a medical system and is therefore resp...

Page 97: ...an be connected or changed any time as described below regardless of whether the system is powered ON or OFF The Venue has four RS connectors Figure 3 25 Probe Connectors on Venue Front End Door CAUTI...

Page 98: ...ing the probe a Do a visual check of the probe pins and system sockets b Remove any dust or foam rests from the probe pins c Verify the probe and the probe cable for any visual damage CAUTION Do not a...

Page 99: ...obe ensure that the connector locking handle is positioned to the right 3 Align the connector with the probe port and carefully push into place 4 Push the connector locking handle to the left to secur...

Page 100: ...ion describes how to configure the Venue 3 6 2 Venue Configuration 3 6 2 1 EZ Config Setup Wizard The Venue Configuration Wizard enables the user to easily configure the system after SW installation 1...

Page 101: ...installation The settings screen appears The Local tab opens by default 4 Wait while system initializes the setup wizard Note For software version 302 x x the initialization duration might take up to...

Page 102: ...ed details and tap Save Then proceed to the next tab 6 On the Network tab define the required settings and tap Save Then proceed to the next tab 7 On the Thermal Printer tab configure the printer and...

Page 103: ...REVISION 6 VENUE SERVICE MANUAL 3 30 Section 3 6 Configuration P R E L I M I N A R Y the printer is already configured skip this step and proceed to the next one Figure 3 32 Venue Installation Wizard...

Page 104: ...pter 3 System Setup 3 31 P R E L I M I N A R Y 8 On the InSite tab define the required settings and tap Submit Changes Make sure you fill all mandatory fields highlighted in bold Then proceed to the n...

Page 105: ...tems list by moving the On Off sliders to On Then select each Connectivity item to define its properties Tap Apply and then Save Then proceed to the next tab 10 On the OptionKey tab enter the key stri...

Page 106: ...hen proceed to the next tab 12 On the About tab view all settings defined on each tab and click Save to save the entire set of defined settings Then click Exit to exit the Setup Wizard NOTE To reiniti...

Page 107: ...Accessing System Configuration Settings All settings configured by the EZ Config wizard can be modified via System Configuration Settings To access these settings perform the following steps 1 On the...

Page 108: ...771498 100 REVISION 6 VENUE SERVICE MANUAL Chapter 3 System Setup 3 35 P R E L I M I N A R Y 3 From the Config side menu select System Settings The System Settings screen is displayed Figure 3 40 Syst...

Page 109: ...DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL 3 36 Section 3 6 Configuration P R E L I M I N A R Y 3 6 2 3 Enter Location Figure 3 41 Enter Location...

Page 110: ...tart The 24 first characters of this name are displayed on the scanning screen s title bar All 64 are displayed on the image properties on saved images 2 Select the Department field and type the name...

Page 111: ...Format see 3 in the figure 24 the 24 hour format is used 12 the 12 AM PM hour format is used 4 Adjust the date see 1 in the figure New date is displayed 5 Adjust the time see 2 in the figure New time...

Page 112: ...98 100 REVISION 6 VENUE SERVICE MANUAL Chapter 3 System Setup 3 39 P R E L I M I N A R Y 3 6 2 5 Select Language for User Interface and Online Manuals Figure 3 43 Select Language Units and Controls So...

Page 113: ...referred language for the online manual The selected language will be used as soon as the unit has been restarted 4 Select the Secondary check box to enable the multi language keyboard characters supp...

Page 114: ...1 SW Recovery In cases of abnormal system behavior as a part of troubleshooting process it is recommended to reload the system software To reload the software without affecting patient data tap the SW...

Page 115: ...n or a combination of software options The Software Option Key is specific for each unit NOTE There may be more than one Software Option Key in use depending on the installed options 3 6 5 2 To Instal...

Page 116: ...ossible connection methods as outlined below 3 7 2 Stand alone Venue No network connection needed 3 7 3 Sneaker Net Environment No network connection needed Use removable media to move data from the V...

Page 117: ...rk printer you should use default delivery settings 3 8 1 Introduction To be able to use the network functions when connected to a hospital network the Venue must have a proper network address Before...

Page 118: ...NNN is the scanner s serial number AE Title Venue Port No Default port number 104 2 Server Config Servers List of servers Buttons Use the buttons to Add Modify or Remove servers 3 Remote Path Used for...

Page 119: ...Log on as adm 2 Tap Settings Config Connectivity TCPIP 3 Tap the Network Settings button The Network Connections window opens 4 Open the Control Panel and go to Network and Internet Network and Sharin...

Page 120: ...ecurity type field select the type of security used by the hidden wireless network Some routers may name this authentication method Depending on the security type you select Windows 10 may or may not...

Page 121: ...rivacy at risk because skilled professionals can intercept this search for the hidden network NOTE If you select the Connect even if the network is not broadcasting check box Windows 10 searches for t...

Page 122: ...6 VENUE SERVICE MANUAL Chapter 3 System Setup 3 49 P R E L I M I N A R Y complete the setup If you are in the range of the hidden WiFi your Windows 10 device automatically connects to it Figure 3 49...

Page 123: ...I M I N A R Y 3 8 4 Changing the AE Title and or Port Number Port No Figure 3 50 AE Title and Port No 1 To change AE Title and or Port No edit the respective fields 2 Select Save settings to store you...

Page 124: ...ironment This procedure is required for every new wireless network NOTE The WiFi configuration is available only if wireless adapter is connected to the system Do not use any type of wireless network...

Page 125: ...nnected to the USB port located on the rear side of the cockpit monitor NOTE It is possible to connect the Wireless Network adapter both when the system is powered ON and after it is powered OFF All s...

Page 126: ...Setup 3 53 P R E L I M I N A R Y Section 3 10 InSite Default Machine Contact Setup 3 10 1 Time Required 5 min 3 10 2 Preparations No dongle or service password needed 3 10 3 Setup Procedure 1 On the m...

Page 127: ...3 6 Configuration P R E L I M I N A R Y 2 From the InSite ExC menu select Request for Service 3 In the Contact GE Service Browser window open the Users tab Figure 3 53 InSite ExC Main Menu Request fo...

Page 128: ...r 5 Enter person s contact information Red asterisk indicates required field Then tap Add User 6 Select the check box for the person who will be the Default Contact Tap Set Default Machine Contact Fig...

Page 129: ...6 Section 3 6 Configuration P R E L I M I N A R Y 7 Tap Set Default Machine Contact 8 The Default Machine Contact will be indicated in red text The setup is complete Figure 3 58 Users Tab Default Mach...

Page 130: ...Installation Acceptance Test Criteria A Venue is ready for use after the system has been configured successfully in accordance with the information provided in Chapter 3 System Setup this chapter 3 11...

Page 131: ...des the General Procedures and the Functional Checks General Procedures is a collection of commonly used procedures that are available by cross references from other parts of this manual Functional Ch...

Page 132: ...ly in place The covers are required for safe operation good Ultrasound system performance and cooling purposes WARNING WARNING Energy Control and Power Lockout for Venue When servicing parts of the Ul...

Page 133: ...t disconnect any USB mass storage device from the system unless a Software Installation procedure is required and the appropriate software installation storage device is connected DANGER ALWAYS CONNEC...

Page 134: ...g label on the rear of the Venue ultrasound scanner 5 Connect the AC Power Cable female plug to the Power Inlet AC IN at the rear of the unit see Figure 4 1 6 Screw the plastic cable clamp into the ho...

Page 135: ...wn after the application is loaded b The Back End Processor and the rest of the scanner starts with the sequence listed in the next steps 1 Back End Processor is turned ON and starts to load the softw...

Page 136: ...Dongle Select this button when you want to exit to the Windows Desktop NOTE If you need to restart Venue when logged on to the Windows Desktop ensure that you do a complete power down Shut Down This...

Page 137: ...wer switch located on the rear lower part of the system to the OFF position Wait for 10 seconds and re set the power switch to the ON position Wait for about 15 Sec before the System is operational Ca...

Page 138: ...screen In order to disable enable the functionality of the On Off physical button open Settings Config Imaging Global Settings and check or uncheck the Disable physical power button box When the physi...

Page 139: ...VENUE SERVICE MANUAL Chapter 4 General Procedures and Functional Checks 4 9 P R E L I M I N A R Y To switch off the system tap Settings Off This action will trigger the display of the System Exit wind...

Page 140: ...ess to the setup menus Example To select a printer As default no password has been set for USR Just type the name USR and select Login If you log on as ADM you will have access to all setup menus and...

Page 141: ...can be adjusted To tilt the cockpit monitor use the lever handle and adjust the cockpit monitor to the desired angle 4 2 8 Moving and Transporting the Venue Ultrasound Scanner 4 2 8 1 The Casters Whee...

Page 142: ...ork 4 Secure the unit s power cable 5 Place all probes in the probe holders Ensure that the probe cables do not protrude from the unit or interfere with the wheels 6 Ensure that no loose items are lef...

Page 143: ...n transporting the Venue by vehicle In addition to the precautions listed earlier follow the steps below Center the cockpit monitor and move it to its down position Disconnect all probes and secure th...

Page 144: ...e operates as intended The functional checks may also be used during troubleshooting 4 3 2 Performance Checks 4 3 2 1 Test Phantoms The use of test phantoms is only recommended if required by your fac...

Page 145: ...Depth control to adjust the range to be imaged Use the Modes to toggle between the scanning system s main modes Use the Frequency control move to lower frequency to increase penetration Press to activ...

Page 146: ...atomical M Mode can also be used with previously acquired digitally stored 2D images More than one heart cycle should be stored if performing M Mode in post processing 2 Position the cursor over the r...

Page 147: ...3 4 5 1 Color 2D 1 From an optimized 2D image press Color 2 Position the ROI frame over the area to be examined 3 Press Select The instruction Size should be highlighted 4 Adjust the dimension of the...

Page 148: ...limit settings Adjust the Compress setting to balance the effect of stronger and weaker echoes and obtain the desired intensity display Adjust Frequency to optimize flow display Higher setting will im...

Page 149: ...and low Pulse Repetition Frequency PRF When the Doppler PRF is raised beyond a certain limit more than one Doppler gate is displayed on the screen Press Auto on the Control panel to activate Automati...

Page 150: ...rator Panel A list of the connected probes will pop up on the screen 2 Select the desired probe An application menu for the desired probe is listed on the screen 3 Browse to the desired application Pr...

Page 151: ...eze to freeze the cineloop 4 Scroll through the acquisition and find the sequence of interest 5 Adjust Cycle select to move from heart beat to heart beat and select the heart cycle of interest 6 Adjus...

Page 152: ...udio Check 4 3 8 1 Introduction The system default settings is to play audible feedback while pressing any action key i e storing an image changing mode typing text etc This functionality can be disab...

Page 153: ...lled and hooked up at the Windows level 1 On the Venue application main screen tap Settings Config Connectivity Additional Outputs Figure 4 12 In the Printer Setup section verify that the selected pri...

Page 154: ...In the Sony UP D711MD window select Printer Properties Printer Window Figure 4 14 In the Sony UP D711MD Properties window tap Print Test Page this will send a print to the printer bypassing all of th...

Page 155: ...ll see an incomplete Test Page printed out This is normal 4 3 9 3 ECG Checks The system automatically detects the ECG module once connected The indication for the user will be displayed in the Periphe...

Page 156: ...nsure that you do not apply too much force to move the cockpit and that the movement is smooth During the movement up and down make sure the image displayed on the monitor does not present any disturb...

Page 157: ...k handle it is located under the cockpit on the front of the Venue system to release the tilt locking mechanism Make sure the locking mechanism is locked when reaching 45o The tilt locking mechanism i...

Page 158: ...ock 1 Push down on the upper grey lever to unlock the brake and swivel on the casters Push and pull the unit right left backwards and forwards Ensure that the wheels move freely in all directions Chec...

Page 159: ...4 29 P R E L I M I N A R Y 4 3 11 Electrical Tests 4 3 11 1 Ground Continuity Test 1 Using multimeter ensure there is a ground continuity between the ground prong of the IPP power connector and the ar...

Page 160: ...heory 5 1 P R E L I M I N A R Y Chapter 5 Venue Components and Function Theory Section 5 1 Overview 5 1 1 Purpose of Chapter 5 This chapter explains Venue system concepts component arrangement and sub...

Page 161: ...ile and depending upon the installed software can be used for various imaging modes These include 2D Gray Scale and 2D Color Flow imaging M Mode Gray Scale imaging Color M Mode Doppler Different combi...

Page 162: ...ectronics The Venue system internal electronics are divided into three Front End Unit see page 5 18 Back End Processor see page 5 22 System Power Distribution see page 5 26 Venue system internal elect...

Page 163: ...y unit and batteries compartment Section 5 3 Options For a list of the available options see Optional Peripherals on page 9 18 For descriptions of the options see the User Manual For Installation inst...

Page 164: ...ways Connection via a Local Area Network LAN set up for this special use only Connection via the hospital network Local Area Network LAN or Wide Area Network WAN The database on the review station is...

Page 165: ...n InSite ExC is your direct link with a GE Online Service Engineer or Applications Support Engineer or a Request for Service via the InSite ExC link at the bottom of the display screen 5 5 2 InSite Ex...

Page 166: ...the customer to generate a Service or Apps request Connect to GE Increase polling rate Click this icon to increase the contact Poll Rate from 15 minutes to 15 seconds The increased polling rate contin...

Page 167: ...patch directly to GE Service after you fill in the following information Items with a red asterisk Problem type Problem area Problem description Send 3 After you have completed filling in all of this...

Page 168: ...The Venue system automatically submits a Request for Service once a software update package was downloaded successfully Applicable for software version 302 X X only All RFS requests can be monitored u...

Page 169: ...our system but cannot perform any service related functions depending on whether InSite has connected or not connected There are two Non Disruptive states If you see a black and white icon InSite ExC...

Page 170: ...0 IoT 64 bit The cockpit monitor module contains the following elements Four probe holders and gel cup holder Power ON OFF touch button SLEEP mode indicator AC Power LED Indicator Speaker Auxiliary di...

Page 171: ...power on Orange indicates system power off The On Off touch button is sensitive to A bare finger A Latex gloved hand A Latex gloved hand with ultrasound gel 5 6 1 3 Sleep Mode Indicator The Sleep Mod...

Page 172: ...r Possible States System State LED Indicators No Power AC main power switch is off or power cord is disconnected batteries are drained or disconnected System is Off AC main power switch is on and powe...

Page 173: ...h scrolling browsing and scaling of the display as required To facilitate comfortable positioning for the operator the height of the Main Display can be adjusted as required For optimal viewing the sc...

Page 174: ...LCD Controller and Touch Module Configuration Interface 5 6 1 9 1 Input Video Interface A Display Port standard receptacle connector is used for carrying video signals and preparatory none video signa...

Page 175: ...Battery Status Indication When the battery is charging If the AC power cable is plugged in the AC mains power indicator is lit green and the battery is being charged until reaching full charge The bac...

Page 176: ...enue ultrasound unit interface panel showing external peripheral accessory connectors 1 Ethernet LAN connector 1000 Base TX Ethernet IEEE 802 3 3kV insulation 2 Dual USB 3 0 connector not insulated 3...

Page 177: ...following modules Front End Power Supply T FEPS See Front End Power Supply T FEPS on page 5 20 Control Front End T CFE Module includes four Transmit and Receive Modules TRX32 See Front End T CFE Boar...

Page 178: ...Module The PSB is connected to the T CFE via 3 connectors The Interface between the T CFE and the PSB modules consists of the following Dedicated I2C channel for Board ID VPD Control Signals I2C chan...

Page 179: ...ansmit receive all the channels data signals to from the four TRX32 boards assembled on the T CFE board In addition it contains all the control and status lines required for HV MUX Setup Probe selecti...

Page 180: ...mechanical and electrical interface for 4 RS probes The T PSB supports the following Probes with up to 128 elements directly 192 element probes using on board HV multiplexers The PSB receives all the...

Page 181: ...Back End Processing BEP Module Solid State Hard Drive SSD 5 9 2 Back End Processing BEP Module The Main BEP module is the System Computer responsible for operating the Venue Scanner application under...

Page 182: ...signals between the BEP and the Front End BEP Interface Panel this panel allows external connections of multiple interfaces by extension cables that connects the internal ports to the IP panel The BEP...

Page 183: ...DIRECTION 5771498 100 REVISION 5 VENUE SERVICE MANUAL 5 24 Section 5 9 Back End Processor P R E L I M I N A R Y 5 9 3 BEP Block Diagram Figure 5 18 BEP Block Diagram...

Page 184: ...he boards in the system and redirect them to their relevant destinations The Back End Board BIB module responsible for the scanner Interfaces comprises three main Interfaces Back End mITX standard via...

Page 185: ...supplies Power to the system when AC power line is not present or in the event of AC power failure 12 16 8V Input filter unit IPP before PSU for filtering AC line noises 5 10 2 AC Distribution The PS...

Page 186: ...d battery power 12v 16 8v The Embedded Controller SOM is responsible for the following functions Control the power up and wake up sequences Battery Monitoring and Battery Charger control Interface to...

Page 187: ...rature is between 0 C and 40 C 32 F and 104 F and discharge the batteries between 20 C and 50 C 4 F and 122 F Do not short circuit the battery by directly connecting the battery terminals with metal o...

Page 188: ...FE Fan located on the top of the Front End module MPB Blower Fan located on the on the rear side of the system behind the MPB module All fans are controlled and have variable speed The cooling require...

Page 189: ...ible for the acquisition of the ECG analog signal inputs 5 12 2 External Peripherals 5 12 2 1 Wi Fi Adapter The Wi Fi adapter allows the user to connect the system to network wireless The Wi Fi adapte...

Page 190: ...come from its integration with iLinq The following sections contain a brief introduction of iLinq s features 5 13 4 1 Service Desktop The Service Platform and other Service software use the iLinq web...

Page 191: ...m 5 13 5 3 Access Security The service interface has different access and security user levels Each user is only granted access to the tools that are authorized for their use NOTE A Service Dongle par...

Page 192: ...33 P R E L I M I N A R Y Every access request whether successful or not will be logged into a service access log that is viewable to authorized users Related information Data Management on page 4 10...

Page 193: ...enue Section 6 2 Power Supply Adjustments There are no adjustments on the power supply The DC Power is self regulated If a voltage is outside the specified range it means that something is wrong eithe...

Page 194: ...creased decreased in the following axes red arrow shows location of the controlling screw per axis 6 4 2 Tilt Resistance Adjustments 1 At the back of the cockpit disconnect the Wi Fi dongle if exists...

Page 195: ...Chapter 6 Service Adjustments 6 3 P R E L I M I N A R Y 3 Using 8mm Allen key and 19mm open wrench release tighten the Allen screw NOTE Ensure there is no tilt autonomic movement Figure 6 22 Cockpit R...

Page 196: ...1 Tighten Release the 8mm socket at the back of NOTE Ensure there is no pan swivel autonomic movement 6 4 4 Arm Vertical Movement Adjustments 1 Move the cockpit to one side until the Alen screw insid...

Page 197: ...e there is no arm autonomic movement 6 4 5 Arm Rotate Resistance Adjustments 1 At the back of the system release the nut under the riser cover 2 Tighten or release the 4mm Allen screw to adjust the re...

Page 198: ...nostics Troubleshooting Section 7 1 Overview 7 1 1 Purpose of Chapter 7 This chapter describes how to setup and run the tools and software that help maintain image quality and system operation Very ba...

Page 199: ...F CAUSING DEATH ARE PRESENT IN THIS EQUIPMENT USE EXTREME CAUTION WHEN HANDLING TESTING AND ADJUSTING If the covers are removed from an operating Venue some metal surfaces may be warm enough to pose a...

Page 200: ...The Venue Companion is a Windows based application that provides visual guides for customer replaceable units CRU This application is distributed as part of the software installation media 7 3 2 Syst...

Page 201: ...2 Service Safety Considerations P R E L I M I N A R Y Press and hold ON OFF Standby buttons simultaneously For R2 systems with software version 302 X X On the Home screen tap Settings Service Diag Fi...

Page 202: ...elect Config F2 About screen Applications Software Application Software revision Software medium s part number System Software System Software revision Software medium s part number 7 4 4 Collect a Tr...

Page 203: ...on 302 X X On the Home screen tap Settings Log Logs NOTE You MUST select one of the available devices as the destination device if it is to be different than the default Export directory on the hard d...

Page 204: ...creation of a log file based on a selected bookmark or for a user configurable time frame Different type of information can be selected to be part of the log file The disk drive installed in the Venue...

Page 205: ...DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL 7 8 Section 7 4 Gathering Troubleshooting Data P R E L I M I N A R Y 7 4 4 2 Figure 7 32 Exiting Service Diagnostic Tests...

Page 206: ...on as possible about the conditions when the noise appear Is the noise present all the time after some time of use After how long time at special times of the day or night When at all locations in the...

Page 207: ...al occasions 7 5 4 4 Self generated Noise Generated inside the Ultrasound system Example Color Noise in the near field Self generated noise will not change if you touch the scanner or the probe Self g...

Page 208: ...shared with any other circuit and a full size ground wire from the distribution panel to the Ultrasound outlet Sites with a mains power system without a defined Neutral The dedicated line shall consis...

Page 209: ...oting 7 6 1 Purpose of this Section In this section you will find Audio troubleshooting procedures and hints to be used if there is no Doppler or system sound 7 6 2 Audio Troubleshooting Procedure 1 I...

Page 210: ...00 REVISION 6 VENUE SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 13 P R E L I M I N A R Y 3 Enter CW mode and select volume menu right side increase the volume 4 Exit to windows Figure 7 34...

Page 211: ...E L I M I N A R Y 2 In the windows screen 1 Tap the speaker icon move the slider to 100 make sure it is not muted 2 Tap and hold the speaker icon and select playback devices make sure the default dev...

Page 212: ...SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 15 P R E L I M I N A R Y 3 Tap and hold the default device and select test 4 While the system is turned on disconnect and reconnect the cockpit b...

Page 213: ...not Recognized Procedure 1 Reseat the PSB TCFE and TRX boards according to Venue service manual 2 If the issue is not resolved this may be a probe issue In this case perform the following steps for e...

Page 214: ...This section manages the most common scenarios where the system does not startup 7 8 2 Contents in this Section Cockpit ON OFF Button turns orange BEP BIOS Jumper Setup 7 52 BEP Fan does not Spin ON...

Page 215: ...m The system does not turn ON cockpit on off button turns green for a few seconds and then turns orange Suggested solution Reset BEP BIOS jumper 1 Reset BIOS by changing the jumper position as shown i...

Page 216: ...ion Symptom T CFE and T FEPS LEDs turn On and then off BEP fan does not spin The On Off button on the cockpit turns green for a second and then turns orange Suggested solution Reset MPB to BIB connect...

Page 217: ...l system behavior occurred Suggested solution Reset front end connections 1 Inspect the BIB blue cable if damaged replace the BIB board 2 Reset the BIB blue cable on the T CFE side disconnect and re c...

Page 218: ...play LEDs are off when turning on the system Suggested solution Reset front end connections 1 If fans are spinning a Reset cockpit cables b Reset MPB c Replace cockpit 2 If fans are NOT spinning a Tur...

Page 219: ...nd By Mode verify T_CFE Board Revision Symptom System Stuck in Stand By Mode Suggested solution verify T_CFE board revision 1 Verify T CFE board P N 5764675_X revision X is 5 or higher the label is fo...

Page 220: ...ify if there is a real failure of the MPB perform the following procedure 7 9 2 MPB Diagnostics Procedure Troubleshooting 3 When running MPB Diagnostics if the following screen appears indicating the...

Page 221: ...t procedures for Venue system parts as outlined below NOTE The illustrations provided in this chapter are for illustration purposes only and are subject to change without notice 8 1 2 Visual Guide The...

Page 222: ...nue ultrasound unit as described in Power Shut Down 2 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 2 1 4 Basic Storage Basket Large Storage Bas...

Page 223: ...described in Power Shut Down 2 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 2 2 4 Power Cable Holder Removal Procedure 1 Move the Power Cable...

Page 224: ...Item Label Item 1 Cover MPB Door 10 Cover Front Base FRU 2 Cover Mid Thermal Baffle FRU 11 Cover Left Side Base FRU 3 Cover Scanner Door FRU 12 Cover Right Side Base FRU 4 Cover Right Side eTower FRU...

Page 225: ...CTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 8 Halo Handle FRU 17 Cover Left Base Bottom FRU 9 Cover Printer Insert FRU 18 Cover Cockpit Rear Cables CRU Table 8 2 Venue Ultr...

Page 226: ...ement cover it is necessary to fit these in the appropriate positions Refer to the illustrations provided in the specific Cover Replacement procedure DANGER ELECTRICAL HAZARDS EXIST AT SEVERAL POINTS...

Page 227: ...described in Power Shut Down 8 3 2 4 Lower Front eTower Cover Removal Procedure 1 Gripping the recess at the center of the cover pull the lower front eTower cover out towards you to remove it from th...

Page 228: ...rations Shut down the Venue ultrasound unit as described in Power Shut Down 8 3 3 4 Left Side eTower Cover Removal Procedure 1 Remove all accessories 2 Remove the RS Probe Cover 3 Grip the rubber hold...

Page 229: ...ide eTower cover and make sure the eTower pins go inside the dedicated holes on the inner side of the cover 2 Install the RS Probe Cover 3 Install all accessories Figure 8 4 Installing the Left Side e...

Page 230: ...e right side eTower cover removal procedure is identical to the left side eTower cover removal procedure 8 3 4 5 Right Side eTower Cover Installation Procedure 1 The right side eTower cover installati...

Page 231: ...enue ultrasound unit as described in Power Shut Down 8 3 5 4 Mid Thermal Baffle Cover Removal Procedure 1 Grip the cover at its bottom and pull the cover toward you to remove it from the system 8 3 5...

Page 232: ...t Cover Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower cover Right Side eTower Cover and Left Side eTower Cover 3 Remove the Upper eTower Co...

Page 233: ...s screws 2 Install the Upper eTower Cover on the scanner 3 Refit the following covers Lower Front eTower cover Right Side eTower Cover and the Left Side eTower Cover and RS Probe Cover 4 Install all a...

Page 234: ...er Shut Down 2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 3 7 4 Printer Insert Cover Remova...

Page 235: ...on Procedure 1 Install the Printer Insert Cover on the Upper eTower Cover and secure with two Phillips screws 2 Install the Upper eTower Cover on the scanner 3 Install all accessories Upper eTower Fro...

Page 236: ...moval Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower cover RS probe cover Right Side eTower Cover and the Left Side eTower Cover 3 Loosen the four captive screws t...

Page 237: ...lowing covers Lower Front eTower cover Right Side eTower Cover and the Left Side eTower Cover 3 Install all accessories Right Side eTower Cover Installation Procedure Left Side eTower Cover Installati...

Page 238: ...the wheels in the locked position 8 3 9 4 RS Probe Cover Removal Procedure 1 Grip the RS Probe Cover on the recess on its bottom and gently pull it toward you to disconnect it from the holding magnet...

Page 239: ...n 8 3 10 4 Riser Thermal Cover Removal Procedure 1 Remove the following covers Lower Front eTower Right Side and Left Side eTower covers 2 Open the door 3 Hold the cover and pull it down 4 Remove the...

Page 240: ...all accessories 2 Remove the following covers Lower Front eTower RS Probe Cover Right Side and Left Side eTower covers Upper Front eTower and Riser Thermal Cover 3 Remove cockpit monitor 4 Remove Scan...

Page 241: ...cedures 8 21 DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 6 Remove the riser cover Note Ensure the pressure pin is not dropped if exists Figure 8 11 Removing Riser Cover...

Page 242: ...he riser cover 2 Install four rubber cable guides two short guides on the back side and two long guides in the front side on the riser cover Use a flat screwdriver to gently affix the rubber guide in...

Page 243: ...rmal Cover 7 Install all accessories Arm Installation Procedure Cockpit Monitor Installation Procedure Lower Front eTower Cover Installation Procedure Left Side eTower Cover Installation Procedure Rig...

Page 244: ...level surface with the wheels in the locked position 8 4 1 4 Base Module Removal Procedure 1 Remove all accessories 2 Remove the cockpit monitor 3 Remove the following covers Lower Front eTower RS Pro...

Page 245: ...ION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 6 Remove the base module 7 Remove the four caster wheels Figure 8 15 Removing the Base Module Casters Removal Procedure Four screw...

Page 246: ...mounting screws and tighten them 4 Refit the following covers Right Side and Left Side eTower covers Lower Front eTowe and Front Base Cover 5 Install the cockpit monitor 6 Install all accessories Fig...

Page 247: ...Front End Metal Door Removal 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower Left side eTower Right Side eTower and Mid Thermal Baffle Cover 3 Remove full fron...

Page 248: ...System Modules Replacement Procedures DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 6 Slightly lift and slide the door upwards until completely removed Figure 8 17 Remov...

Page 249: ...9 DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 7 Remove the eight holding screws to separate the front end metal door from the scanner door cover Figure 8 18 Separating...

Page 250: ...n cable 6 Close the FE door and close the two fastening latches on the left side of the Front End door assembly 7 Tighten the captive screw 8 Refit the following covers Mid Thermal Baffle Cover Lower...

Page 251: ...a level surface with the wheels in the locked position 8 4 3 4 Full Front End Removal 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower Left side eTower and Righ...

Page 252: ...NUE SERVICE MANUAL P R E L I M I N A R Y 4 Disconnect the following cables MPB to T FEPS BIB to T CFE Blue cable Fan cable 5 While supporting the FE module with one hand unscrew the four Phillips scre...

Page 253: ...with cable clips Fan cable 5 Close the FE door and close the two fastening latches on the left side of the Front End door assembly Figure 8 19 6 Tighten the captive screw 7 Refit the following covers...

Page 254: ...4 SSD Module Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower Left side eTower Right Side eTower 3 Open the two fastening latches on the left...

Page 255: ...he Front End door assembly 3 Tighten the captive screw 4 Refit the following covers Lower Front eTower left side eTower right side eTower and RS Probe Cover Ins 5 Install all accessories 6 Turn ON pow...

Page 256: ...trasound unit as described in Power Shut Down 2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8...

Page 257: ...it rear cables cover if exists see below 4 Disconnect the following cables connected to cockpit monitor BE TO COCKPIT CABLE MPB TO COCKPIT CABLE use flat head screwdriver Figure 8 23 Cockpit Monitor S...

Page 258: ...he dedicated hole and press to release the securing spring The click is sounded indicating the securing spring is released 7 Lift and remove the cockpit monitor Figure 8 25 Cockpit Ground Cables Remov...

Page 259: ...e 1 Mount the cockpit monitor on the mounting hooks and secure with four Phillips screws 2 Reconnect the cockpit monitor cables BE TO COCKPIT CABLE MPB TO COCKPIT CABLE Ground cables x2 Rear cable cov...

Page 260: ...anding securely on a level surface with the wheels in the locked position 8 4 6 4 PSU Module Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower...

Page 261: ...ve screw at the bottom of the PSU and pull the PSU out gently 8 4 6 5 PSU Module Installation Procedure 1 Slide the new PSU into position 2 Secure the captive screw at the bottom of the PSU 3 Install...

Page 262: ...Side eTower Cover Left Side eTower Cover and RS Probe Cover 6 Install all accessories MPB Door Cover Installation Procedure Right Side eTower Cover Installation Procedure Left Side eTower Cover Instal...

Page 263: ...s in the locked position 8 4 7 4 Battery Module Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower cover Right Side eTower Cover Left Side eTowe...

Page 264: ...crews for each battery 3 Refit the following covers Lower Front eTower cover Right Side eTower Cover Left Side eTower Cover MPB Door Cover and RS Probe Cover 4 Install all accessories MPB Door Cover I...

Page 265: ...s standing securely on a level surface with the wheels in the locked position 8 4 8 4 BE Module Removal Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower cover Left S...

Page 266: ...ly and loosen the captive screw Then carefully open the door 6 Secure the door with door support 7 Disconnect the following cables connected to the BE module MPB to BIB MPB to BEP white USB cable Cock...

Page 267: ...L I M I N A R Y 9 Disconnect the BIB blue cable from the CFE plastic cable guide Disconnect the MPB to BIB cable 10 Loosen the two captive screws on both sides of the BE module 11 Gently slide out th...

Page 268: ...EP white Cockpit to BEP split cable green and black Printer cable yellow Use the following diagram for reference 5 Reconnect the BIB blue cable to the CFE plastic cable guide 6 Reconnect the MPB to BE...

Page 269: ...ories 13 Turn ON power to the system Left Side eTower Cover Installation Procedure Right Side eTower Cover Installation Procedure Lower Front eTower Cover Installation Procedure RS Probe Cover Install...

Page 270: ...wheels in the locked position 8 4 9 4 BEP PCB Removal Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower Left side eTower and Right Side eTower 3 Perform BE module rem...

Page 271: ...of the bracket 6 Release the following cable connectors SATA DATA BE TO SSD A red cable press the latch and pull the cable connector out SSD PWR cable Gently pull the connector towards you 7 Disconne...

Page 272: ...PCB Release four screws in the BEP PCB corners using Phillips screwdriver Note If the MiTx to BIB cable is preventing easy removal of the BEP PCB disconnect the MiTx to BIB cable connector latch see...

Page 273: ...cedure 1 Reconnect the BEP PCB using four screws in the PCB corners using Phillips screwdriver 2 Reconnect the following cables at the back of the BE Two USB connectors Network cable Display port 3 Re...

Page 274: ...ENUE SERVICE MANUAL P R E L I M I N A R Y SSD PWR cable 4 Reconnect the supporting bracket using an appropriate Phillips screwdriver Two captive screws Two screws at the bottom of the bracket Figure 8...

Page 275: ...eTower Front Cover 9 Install all accessories 10 Turn ON power to the system Figure 8 45 BE Omega Bracket Repositioning MPB Module Installation Procedure MPB Front Metal Door Installation Procedure Up...

Page 276: ...eels in the locked position 8 4 10 4 MPB Module Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe cover Lower Front eTower cover Right Side eTower cover Left Side eTowe...

Page 277: ...ION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 6 Release the captive screw at the bottom center of the MPB 7 Pull the MPB halfway out 8 Tilt the top toward you and pull the MPB out all the way Figur...

Page 278: ...the rails until the docking of the MPB module 2 Tighten the two mounting screws to secure the MPB module 3 Return the PSU module 4 Return the battery packs 5 Refit the following covers MPB Door Cover...

Page 279: ...SERVICE MANUAL P R E L I M I N A R Y 8 Check PM_FW 9 If PM_FW is not up to date the Comment section states expected version is 3 xx perform software reload 10 Figure 8 48 Check if PM_FW is up to date...

Page 280: ...essories 2 Remove the following covers RS Probe Cover Lower Front eTower Left side eTower and Right Side eTower 3 Open and remove the full Front End door assembly and place it on a stable surface 4 Re...

Page 281: ...6 VENUE SERVICE MANUAL P R E L I M I N A R Y 5 Loosen and remove the two retaining screws and remove the TRx box 6 Remove the T TRx 32 modules 7 Loosen and remove the six screws Figure 8 49 Removing...

Page 282: ...he TTRX BOX and secure with two screws previously removed 4 Install the T FEPS module 5 Install the full Front End assembly 6 Refit the following covers Lower Front eTower Left side eTower Right Side...

Page 283: ...5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 7 Install all accessories 8 Turn ON power to the system Accessories Replacement Procedures Perform the checks listed in T CFE cFront...

Page 284: ...side eTower and Right Side eTower 3 Remove the T FEPS module 4 Open and remove the full Front End assembly and place it on your laps in order not to damage the probe locker levers 5 Remove the T CFE...

Page 285: ...wing covers Lower Front eTower Left side eTower Right Side eTower and RS Probe Cover 6 Install all accessories 7 Turn ON power to the system Plastic Cable Guide for CFE Removal Procedure T CFE Module...

Page 286: ...re the system is standing securely on a level surface with the wheels in the locked position 8 5 1 4 BIB Board Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe cover L...

Page 287: ...t 5 Disconnect BIB TO BE SPLIT CABLE connector that connects the BIB board to the BE module 6 Disconnect the PCIe cable connector that connects the BIB board to the BE module Figure 8 52 Removing BE S...

Page 288: ...e two fastening latches on the left side of the Front End door assembly 6 Tighten the captive screw 7 Refit the following covers Lower Front eTower Left side eTower Right Side eTower and RS Probe cove...

Page 289: ...n the locked position 8 5 2 4 CFE Board Release Arm Removal Procedure The lever shaped CFE Release Arm is located on the PSB Board 1 Remove all accessories 2 Remove the following covers RS Probe Cover...

Page 290: ...Install the Front End door assembly 6 Refit the following covers Lower Front eTower Left side eTower and Right Side eTower 7 Install all accessories 8 Turn ON power to the system Figure 8 57 Removing...

Page 291: ...cedure 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower Left side eTower Right Side eTower 3 Open the two fastening latches on the left side of the FE door asse...

Page 292: ...3 Reconnect the cables previously disconnected 4 Close the Front End door and fasten securely with the two latches 5 Refit the following covers Lower Front eTower Left side eTower and Right Side eTow...

Page 293: ...covers RS Probe Cover Lower Front eTower Left side eTower and Right Side eTower 3 Open the two fastening latches on the left side of the FE door assembly and loosen the captive screw Then carefully op...

Page 294: ...reviously disconnected 3 Close the Front End door and fasten securely with the two latches 4 Refit the following covers Lower Front eTower Left side eTower Right Side eTower and RS Probe Cover 5 Insta...

Page 295: ...nd Right Side eTower 3 Open the two fastening latches on the left side of the FE door assembly and loosen the captive screw Then carefully open the door 4 Remove the T TRx Box 5 Secure the door by cli...

Page 296: ...ay of a white plastic holder at each end of the T TRx module These are opened by gently pulling them outwards to tilt the holder away from the edge of the module as shown in Figure 8 60 8 Open each wh...

Page 297: ...the cables previously disconnected 6 Close the Front End door fasten securely with the two latches and tighten the captive screw 7 Refit the following covers Lower Front eTower Left side eTower Right...

Page 298: ...e Venue ultrasound unit as described in Power Shut Down 2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked p...

Page 299: ...lation Procedure 1 Install the new MPB Front Metal Cover and tighten the four captive screws 2 Refit the Lower Front eTower cover 3 Install all accessories 4 Turn ON power to the system Lower Front eT...

Page 300: ...t Down 2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 6 2 4 MPB Guide L R Removal Procedure 1...

Page 301: ...8 6 2 5 MPB Guide L R Installation Procedure 1 Attach the MP Guide L R 2 Tighten the two mounting screws to secure the guide to the receptical box 3 Install the MPB Module Figure 8 64 MPB Guide L R Mo...

Page 302: ...wer Cover and RS Probe cover 5 Install all accessories 6 Turn ON power to the system MPB Door Cover Installation Procedure Left Side eTower Cover Installation Procedure Right Side eTower Cover Install...

Page 303: ...2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 6 3 4 Halo Handle Removal Procedure 1 Remove...

Page 304: ...Remove the Halo Handle mounting screws four x4 front screws and four x4 side screws two from each side 7 Lift the handle upwards and remove it Figure 8 66 Removing the Halo Handle Mounting Screws Fig...

Page 305: ...and tighten them 3 Install the Riser Cover 4 Install the Riser Thermal Cover 5 Install the articulated arm 6 Install the cockpit monitor 7 Install all accessories 8 Turn ON power to the system Riser...

Page 306: ...securely on a level surface with the wheels in the locked position 8 6 4 4 Arm Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower cover Left Sid...

Page 307: ...through the arm axis refer to MPB TO Cockpit Cable Removal Procedure and to BE to Cockpit Cable Removal Procedure 7 Using Allen key screwdriver remove the two securing screws of the arm side cover 8...

Page 308: ...l out the cockpit monitor cables routed inside the riser 11 Disconnect the ground cable from the riser 12 Using Forged Steel Ring Pliers remove the C clip and then the Teflon ring Figure 8 70 Removing...

Page 309: ...cables through the riser opening refer to MPB TO Cockpit Cable Installation Procedure and BE to Cockpit Cable Installation Procedure 4 Route each cockpit monitor cable through the arm side and install...

Page 310: ...11 Turn ON power to the system Riser Thermal Cover Removal Procedure Left Side eTower Cover Removal Procedure Right Side eTower Cover Removal Procedure Lower Front eTower Cover Removal Procedure RS P...

Page 311: ...er Removal Procedure 1 Remove all accessories 2 Remove cockpit monitor 3 Remove the following covers RS Probe cover Lower Front eTower cover Left Side eTower cover Right Side eTower cover Riser Therma...

Page 312: ...00 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 6 Remove the riser cover 7 Remove the Teflon sliding top ring Note Ensure the pressure pin is not dropped if exists Figure 8 75 Removing Riser...

Page 313: ...r 8 6 5 5 Riser Installation Procedure 1 Install new riser on the eTower 2 Apply LOCTITE 243 on the riser securing screws and tighten the screws 3 Install the Teflon sliding top ring 4 Remove the mask...

Page 314: ...ver and RS Probe cover 8 Install the cockpit monitor 9 Install all accessories 10 Turn ON power to the system Arm Installation Procedure Upper eTower Front Cover Installation Procedure Right Side eTow...

Page 315: ...6 4 Casters Removal Procedure 1 Using a 6mm Allen key loosen and remove the screw that secures the wheel shaft in the wheel securing socket 2 Lift the chassis sufficiently to allow the wheel to drop d...

Page 316: ...re the screw is tightened sufficiently to secure the wheel on the chassis Do not over tighten the screw NOTE When installing wheels make sure you install the correct type of the wheel The wheels posit...

Page 317: ...Procedure 1 Remove all accessories 2 Remove the following covers RS Probe cover Lower Front eTower Left side eTower Right Side eTower Riser Thermal Cover 3 Open the two fastening latches on the left...

Page 318: ...ERVICE MANUAL P R E L I M I N A R Y Gently pull down the Plastic Cable Guide for CFE and disconnect the cable connector 6 Loosen and remove the two screws securing the cable guide to CFE board and rem...

Page 319: ...4 Close the FE door and close the two fastening latches on the left side of the Front End door assembly 5 Tighten the captive screw 6 Refit the following covers Lower Front eTower left side eTower Rig...

Page 320: ...vel surface with the wheels in the locked position 8 6 8 4 MPB Blower Fan Removal Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower Left side eTower and Right Side eT...

Page 321: ...dures 8 101 DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 5 Loosen the elastic spring slings that hold the fan and carefully remove the MPB fan Figure 8 86 Removing the M...

Page 322: ...n Procedure 1 Carefully insert the new MPB fan into the fan sleeve 2 Tighten the elastic spring slings to secure the fan in its place 3 Reconnect the MPB to BIB Control Cable to the fan cable connecto...

Page 323: ...ide eTower and RS Probe cover 6 Install all accessories 7 Turn ON power to the system Left Side eTower Cover Installation Procedure Right Side eTower Cover Installation Procedure Lower Front eTower Co...

Page 324: ...e wheels in the locked position 8 6 9 4 IPP Module Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe cover Lower Front eTower Left side eTower Right Side eTower and MPB...

Page 325: ...the three screws that hold the IPP metal frame 8 Using the nose pliers or flat head screwdriver squeeze the white plastic connector holders and push them out to remove the power connector Figure 8 89...

Page 326: ...s Replacement Procedures DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 9 Using the open end wrench disconnect the ground cable 10 Remove the IPP module Figure 8 91 Discon...

Page 327: ...ower Door 8 Refit the following covers MPB Front Metal Door Left side eTower Right Side eTower Lower Front eTower and RS Probe cover 9 Install all accessories 10 Turn ON power to the system MPB Module...

Page 328: ...n the locked position 8 7 1 4 MPB Rear USB Cable Removal Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower Left side eTower and Right Side eTower 3 Remove MPB Front M...

Page 329: ...Connector 8 Using Phillips screwdriver disconnect the two screws that hold the cable connector to the rear panel and remove the cable Figure 8 92 Disconnecting MPB Rear USB Cable To MPB J12 Connector...

Page 330: ...attery pack units 6 Install MPB Front Metal Door 7 Refit the following covers Lower Front eTower left side eTower Right side eTower and RS Probe cover 8 Install all accessories 9 Turn ON power to the...

Page 331: ...els in the locked position 8 7 2 4 MPB TO Cockpit Cable Removal Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower Left side eTower Right Side eTower and Riser Thermal...

Page 332: ...Y b Release the four Phillips screws and remove the rear cable cover 5 Disconnect the MPB TO COCKPIT CABLE from the monitor Disconnect the cable connector use flat head screwdriver Release Ground cabl...

Page 333: ...y raise the arm to its up position b Remove the arm base cover c Using Phillips screwdriver remove the two securing screws of the arm side cover d Slide the right side arm side cover upwards and remov...

Page 334: ...I N A R Y 9 Route the cable and its connector downwards through the arm hole 10 Release plastic clips that hold the cable using a flat screwdriver Three clips in the Front End Two clips between the c...

Page 335: ...not to pull and cause damage to the other cables 13 Open cable cover door Open the captive screw of the cable cover door using a Phillips screwdriver Slide the cable cover door towards you to free the...

Page 336: ...e Installation Procedure 1 Remove ferrite clip from MPB to Cockpit cable 2 Route the MPB to Cockpit cable through the articulated arm so that the length of the exposed cable toward the cockpit is mini...

Page 337: ...orce and route the MPB to Cockpit cable through the tilt axis keeping the routed cable on the most left and the connector facing to the right Make sure it is routed above the tilt release cable wire F...

Page 338: ...able approximately 3 cm below the cable connector 6 Connect the MPB to Cockpit cable to the cockpit and fasten the flat screws 7 Reconnect the ground cable 8 Make sure the MPB to Cockpit cable is rout...

Page 339: ...lower covers 10 Route the MPB to Cockpit cable through the cable cover door and the arm hole 11 Connect the MPB TO COCKPIT CABLE to the monitor Connect the cable connector use flat head screwdriver C...

Page 340: ...ure cable to arm using tie wraps 14 Gently slide the MPB TO COCKPIT cable through the MPB receptacle side towards the front 15 Reconnect J1 and J2 connectors ensuring the connectors arrow is pointing...

Page 341: ...e the cable cover door Slide the cable cover door towards the system and fix the rubber grommet in its position Close the captive screw of the cable cover door using a Phillips screwdriver 18 Install...

Page 342: ...the Wi Fi dongle if exists 25 nstall all accessories 26 Turn ON power to the system MPB Module Installation Procedure Riser Thermal Cover Installation Procedure Left Side eTower Cover Installation Pr...

Page 343: ...to Cockpit Cable Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe Lower Front eTower Left side eTower Right Side eTower Riser Thermal and MPB Door cover 3 Disconnect...

Page 344: ...ollowing steps a Disconnect the Wi Fi adapter b Release the four Phillips screws and remove the rear cable cover c Gently pull out the BE to Cockpit cable from the axis tilt 5 Release the BE to Cockpi...

Page 345: ...side cover d Slide the left side arm side cover upwards and remove it 6 Cut the tie wraps securing the BE to cockpit cable to the arm left side and release the cable from the arm 7 Route the cable an...

Page 346: ...side 9 Disconnect the Cockpit to BEP split cable connected to BE module green USB connector and black display port connector 10 Open cable cover door a Open the captive screw of the cable cover door u...

Page 347: ...r 8 Replacement Procedures 8 127 DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y door with the rubber grommet The cable is released Figure 8 124 Releasing the BE to Cockpit...

Page 348: ...e Installation Procedure 1 Route BE to Cockpit Cable through arm so that the cable length that remains exposed to the cockpit direction is minimum 35cm 2 Remove the lower and then the upper cockpit co...

Page 349: ...trough cockpit tilt axis For easier routing make sure the cable is aligned to the left 4 Connect the BE to Cockpit cable DP connector to the cockpit 5 Reconnect the ground cable 6 Refit the cockpit up...

Page 350: ...use flat head screwdriver Connect Ground cable use Phillips screwdriver 9 Gently slide the BE TO COCKPIT cable through the MPB receptacle side towards the front 10 Reconnect the Cockpit to BEP split c...

Page 351: ...able cover door towards the system and fix the rubber grommet in its position Close the captive screw of the cable cover door using a Phillips screwdriver 13 Install the arm side cover 14 Secure side...

Page 352: ...accessories 20 Turn ON power to the system Riser Thermal Cover Installation Procedure Left Side eTower Cover Installation Procedure Right Side eTower Cover Installation Procedure Lower Front eTower C...

Page 353: ...ed in Power Shut Down 2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 7 4 4 PSU to MPB DC Dock...

Page 354: ...a On each side of each connector use a flat screwdriver to push one of the plastic flaps towards the hole and then push the other flap through the hole to release the connector b From the front side r...

Page 355: ...nsuring the orientation is appropriate by checking the plastic marks are pointing to the correct direction Push the connectors plastic hinges through the designated holes until both hinges are fully i...

Page 356: ...is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 7 5 4 MPB to BE PWR Cable Removal Procedure 1 Remove all accessories 2 Remove the...

Page 357: ...I M I N A R Y 4 Disconnect the two lower cable connectors J4 J5 by sliding them upwards see J connectors map that follows 5 Release the cable from the two plastic cable clips The cable is released Fi...

Page 358: ...owing covers MPB Door Cover Left and Right Side eTower Covers Lower Front eTower Cover and RS Probe cover 6 Install all accessories 7 Turn ON power to the system MPB Module Installation Procedure MPB...

Page 359: ...eels in the locked position 8 7 6 4 ON OFF Switch Cable Removal Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower Left side eTower and Right Side eTower 3 Remove the...

Page 360: ...Disconnect the ON OFF Switch cable connector from the MPB J10 connector Disconnect the ON OFF Switch cable connector with the ON OFF switch from the MPB module by pressing the securing clips on both s...

Page 361: ...e MPB module and tighten the two mounting screws 4 Install the battery pack units 5 Refit the following covers Lower Front eTower left side eTower Right side eTower and RS Probe cover 6 Install all ac...

Page 362: ...face with the wheels in the locked position 8 7 7 4 PCIe Cable Removal Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower Left side eTower and Right Side eTower 3 Remo...

Page 363: ...BE module 3 Refit the following covers Lower Front eTower Left side eTower and Right Side eTower 4 Install all accessories 5 Turn ON power to the system BE Module Installation Procedure Right Side eTo...

Page 364: ...ecurely on a level surface with the wheels in the locked position 8 7 8 4 MPB to BIB Control Cable Removal Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower Left side...

Page 365: ...UAL P R E L I M I N A R Y 6 Disconnect MPB to BIB Control Cable from the fan cable connector 7 Release plastic clips that hold the cable Figure 8 146 Disconnecting Fan Cable Connector Figure 8 147 Rel...

Page 366: ...he blower cable connector 3 Secure the cable with plastic clips 4 Install the MPB module 5 Install the MPB Front Metal Door 6 Install the following covers Lower Front eTower Left side eTower Right Sid...

Page 367: ...th the wheels in the locked position 8 7 9 4 Printer USB Cable Removal Procedure 1 Remove all accessories 2 Remove the following covers RS Probe Cover Lower Front eTower Left side eTower Right Side eT...

Page 368: ...ve screw NOTE Make sure the Peripheral Cable Cover is properly aligned and affixed 4 Install MPB Front Metal Door 5 Refit the following covers RS Probe Cover Lower Front eTower Left side eTower Right...

Page 369: ...s standing securely on a level surface with the wheels in the locked position 8 7 10 4 BEP Cables Removal Procedure 1 Remove all accessories 2 Remove the following covers Lower Front eTower Left side...

Page 370: ...aptive screws Release two screws at the bottom of the bracket 6 Release the following cable connectors SATA DATA BE TO SSD A red cable press the latch and pull the cable connector out SSD PWR cable Ge...

Page 371: ...ble is preventing easy removal of the PCB disconnect the MiTx to BIB cable connector latch see yellow circle in the above figure Gently take out the PCB 9 The cables are now exposed and easy to access...

Page 372: ...WR cable SATA DATA BE to SSD cable MiTx to BIB cable release one connector on the BIB and pull upwards two connectors on the BEP PCB 8 7 10 5 BEP Cables Installation Procedure 1 Reconnect the cable s...

Page 373: ...R Y 2 Route the cables on the BEP bottom bracket as shown below 3 Reconnect the PCB using four screws in the PCB corners using Phillips screwdriver 4 Reconnect the following cables at the back of the...

Page 374: ...E L I M I N A R Y Display port 5 Reconnect the following cable connectors SATA DATA BE TO SSD A red cable SSD PWR cable 6 Reconnect the supporting bracket using an appropriate Phillips screwdriver Tw...

Page 375: ...ing covers RS Probe Cover Lower Front eTower Left side eTower Right Side eTower Upper eTower Front Cover Figure 8 162 Reconnecting BE Supporting Bracket Figure 8 163 BE Omega Bracket Repositioning MPB...

Page 376: ...es DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 11 Install all accessories 12 Turn ON power to the system Accessories Replacement Procedures Perform the checks listed in...

Page 377: ...ltrasound unit as described in Power Shut Down 2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8...

Page 378: ...he MPB TO T FEPS cable edge 3 Connect the following cables to BE module MPB to BIB MPB to BEP white Cockpit to BEP split cable green and black Printer cable yellow 4 Install the MPB module 5 Install t...

Page 379: ...771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 7 Install all accessories 8 Turn ON power to the system Accessories Replacement Procedures Perform the checks listed in Front End Power...

Page 380: ...ired 15 min 8 7 12 3 Preparations 1 Shut down the Venue ultrasound unit as described in Power Shut Down 2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on...

Page 381: ...e cable from the upper plastic cable clips 4 If a printer is used Release the printer plastic cable clips routed upwards Loosen the Phillips screw and remove the Peripheral Cable Cover Figure 8 167 Di...

Page 382: ...ENUE SERVICE MANUAL P R E L I M I N A R Y Disconnect the printer cable yellow USB connector from the Back End assembly 5 Disconnect the MPB to BEP white cable connected to BE module Figure 8 170 Relea...

Page 383: ...clips 4 Install the MPB module 5 Install the MPB Front Metal Door 6 Install the following covers Lower Front eTower Left side eTower Right Side eTower and RS Probe cover 7 Install all accessories 8 T...

Page 384: ...e wheels in the locked position 8 8 1 4 Printer Removal Procedure NOTE Perform these steps only for Venue systems with printer already installed on them For systems with no previously installed printe...

Page 385: ...g Printer 1 Remove the Printer Insert Cover from the Upper eTower Cover 2 Install the Printer Frame Cover and secure with two Phillips screws 3 Remove the metal bracket Use Phillips screwdriver Printe...

Page 386: ...lip the white plastic cable holder and release the two printer cables Printer Power Cable and Printer USB Cable 6 Attach bracket to the Printer and tighten the four Phillips screws the bracket and the...

Page 387: ...MANUAL P R E L I M I N A R Y 7 Insert the Printer Power Cable and Printer USB cable through the dedicated opening 8 Connect the Printer USB Cable and Printer Power Cables to the Printer Figure 8 177 P...

Page 388: ...L I M I N A R Y 9 Attach the Printer to the eTower and secure with two bottom screws 10 Attach the metal cable cover and secure the Phillips screw 11 Install MPB Front Metal Door Figure 8 179 Attachin...

Page 389: ...ter Installation Procedure For Systems with Existing Printer 1 Perform Steps 8 14 of the section Printer Installation Procedure For Systems with No Existing Printer Upper eTower Front Cover Installati...

Page 390: ...ke sure the system is standing securely on a level surface with the wheels in the locked position 8 8 2 4 Wi Fi Adapter Removal Procedure NOTE Perform these steps only for Venue systems with Wi Fi ada...

Page 391: ...ously installed Wi Fi adapter For systems with existing Wi Fi adapter skip this step and proceed to Step 2 2 Connect the L shape USB adapter to the back of the Cockpit 3 Attach the transparent Wi Fi d...

Page 392: ...Installation Procedures DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y Note The barcode device is detected by the system as a COM Port under windows device manager Figure...

Page 393: ...is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 8 3 4 ECG Installation 1 Remove the following covers Lower Front eTower Upper Fron...

Page 394: ...s and 2 washers verify correct orientation 4 Using two Thumb screws install the bracket on the eTower 5 Connect the ECG USB cable to the USB 2 0 port on the BEP above the Green connector 6 Route the E...

Page 395: ...G USB cable 8 Fold the excessive cable and use the cable clips to hold it 9 Attach the metal bracket to close the cable opening 10 Route the USB cable as shown below and connect the ECG USB cable to t...

Page 396: ...the ECG module and tighten the screw to fix the ECG module in place 12 Install the following covers MPB Front Metal Door Upper Front eTower Lower Front eTower with cutout for ECG Figure 8 192 Fixing...

Page 397: ...ower Shut Down 2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 8 4 4 ECG Installation 1 Remove...

Page 398: ...thumbscrews holding the printer bracket and remove the printer with the bracket 4 Remove four screws connecting the printer to the bracket 5 Attach the ECG bracket to the printer bracket with 2 screw...

Page 399: ...printer using four screws 7 Connect ECG USB cable to the USB 2 0 port on the BEP on top of the green connector 8 Route the USB cable through the dedicated opening in the eTower and affix it along wit...

Page 400: ...he metal bracket 10 Connect the USB cable and power cable to the printer and the USB cable to the ECG module 11 Using two Thumb screws install the bracket on the eTower 12 Tighten the screw to fix the...

Page 401: ...able and use the cable clips to hold it 14 Install the following covers MPB Front Metal Door Upper Front eTower Lower Front eTower with cutout for ECG Figure 8 203 Connecting Printer and ECG Cables MP...

Page 402: ...described in Power Shut Down 2 Make sure the On Off power switch is set to Off 3 Make sure the system is standing securely on a level surface with the wheels in the locked position 8 8 5 4 ECG Remova...

Page 403: ...nd remove the ECG 8 8 5 5 ECG Installation Procedure 1 Connect ECG to the USB cable 2 Insert the ECG module all the way in 3 Tighten the ECG securing screw 4 Install the following covers Upper Front e...

Page 404: ...llow the program to make changes to your computer 3 The ImageUSB by PassMark Software main window appears NOTICE Burning of the disk on key media Mem stick is performed using ImageUSB utility download...

Page 405: ...check box 6 In the Step 3 section click Browse and select the image file in BIN format downloaded from the GE Portal 7 In the Step 4 section click Write to start the burning process Figure 8 208 Disk...

Page 406: ...burning process 9 View the burning progress in the Overall progress status bar 10 If the software burning process is successfully completed the Imaging Completed message appears Click OK to close the...

Page 407: ...M I N A R Y 11 If the burning process fails a notification window appears Click OK to close the window Repeat steps 3 9 using different USB media 12 Exit the ImageUSB utility 13 Proceed with the Soft...

Page 408: ...tions in accordance with different software versions 1 Disconnect all probes 2 Plug the Software Installation Media into the USB port located on the Interface Panel of the Venue 3 Turn on the system T...

Page 409: ...A R Y For R2 software version 302 X X the following Warning message appears 5 Click Yes in the Warning message box 6 Automatic installation process runs At the end of the process the following messag...

Page 410: ...on media and tap OK in the message box 9 In the Confirmation dialog box that appears tap Yes to continue the reboot 10 Tap Yes to continue the reboot 11 After the system reboots the Venue setup wizard...

Page 411: ...Y 13 Tap Run wizard to continue the installation 14 Enter the system serial number and tap OK to confirm 15 Tap OK to confirm the serial number Figure 8 223 Venue Installation Wizard Run Wizard Figure...

Page 412: ...regain the initial storage capacity of the media used for burning software downloaded from GE portal 1 On your PC insert the Disk on Key media into any available USB port 2 Browse to the folder that c...

Page 413: ...r 2 Tap Settings and select Config 3 On the Connectivity tab select Dataflow and make sure the USB Storage option is enabled WARNING WARNING WHILE THE SOFTWARE UPGRADE PROCEDURE IS DESIGNED TO PRESERV...

Page 414: ...ANUAL P R E L I M I N A R Y 4 Connect USB memstick to any available USB port in the system 5 From the side menu select Admin and then Backup 6 Select the System Configuration checkbox 7 Select USB HD...

Page 415: ...he backup is complete the following message appears 9 Eject the USB memstick when prompted 8 9 3 2 Backup of Patient Exams 1 Connect an external USB 3 0 hard drive with at least 100GB of free space 2...

Page 416: ...L I M I N A R Y 3 Select the Select All checkbox to select all patients for backup 4 Tap Send Selected 5 In the Copy window select USB Storage and click OK Figure 8 232 Select All Patients for Backup...

Page 417: ...complete proceed to Software Upgrade Procedure 8 9 3 3 Software Upgrade Procedure 1 Disconnect all probes 2 Plug the Venue R2 Software Installation Media into the USB port located on the Interface Pa...

Page 418: ...nstallation media and tap OK in the message box 9 In the Confirmation dialog box that appears tap Yes to continue the reboot 10 Tap Yes to continue the reboot 11 After the system reboots the Venue set...

Page 419: ...to proceed 13 Tap Run wizard to continue the installation 14 The settings screen appears The Local tab opens by default NOTE Due to FPGAs programming as part of the initialization process it might ta...

Page 420: ...ow the procedure displayed in the window If the following errors appear proceed with T CFE replacement 15 Type in the option key and configure the network settings 16 Continue with the EZ Config insta...

Page 421: ...ON 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y check box is NOT selected Figure 8 244 Auto zone increment check box Perform the checks listed in Operating System and Application...

Page 422: ...Application Software Loading Procedures DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 18 Verify that an e label is displayed in Config screen under the About tab as shown...

Page 423: ...the CRM system is updated correctly see Figure 8 246 NOTE This step is applicable only for upgrades that were obtained electronically eDelivery 20 In case the purchased upgrade kit was delivered as a...

Page 424: ...Operating System and Application Software Loading Procedures DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y Figure 8 247 Example of UDI label attached to Software install...

Page 425: ...are Option keys are available prior to commencing the installation process Do not use the Software Option keys from a previous system version At the site perform the following steps before you start t...

Page 426: ...rating system and the Venue application software This excludes preparation time backup etc and performing functional checks following the installation procedures Throughout the installation processes...

Page 427: ...port located on the Interface Panel of the Venue 3 Turn on the system The opening screen is displayed 4 For new installation after SSD replacement tap Format SSD and Install If patient information is...

Page 428: ...media and tap OK in the message box 9 In the Confirmation dialog box that appears tap Yes to continue the reboot 10 Tap Yes to continue the reboot 11 After the system reboots the Venue setup wizard w...

Page 429: ...VISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 13 Tap Run wizard to continue the installation 14 Enter the system serial number and tap OK to confirm Figure 8 253 Venue Installation Wizard Run Wi...

Page 430: ...Tap OK to confirm the serial number 16 The settings screen appears The Local tab opens by default 17 Continue with the EZ Config installation as specified in Section 3 6 2 1 on page 3 27 Figure 8 255...

Page 431: ...placement Procedures 8 211 DIRECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 8 9 6 NOTE The downgrade procedure is only applicable for R1 systems in which the R2 upgrade proc...

Page 432: ...rences or variations in accordance with different software versions The software recovery procedure allows the user to reload the operating system and the application Drive C partition without affecti...

Page 433: ...SION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 4 A pop up window will be displayed notifying the user about the next steps 5 Select OK Windows recovery options will be displayed 6 Select Troublesho...

Page 434: ...ue SW reload 8 The system will reboot automatically and the Venue SW reload window will be displayed 9 Select Reload Venue SW The software reload process will start 10 Once the process is complete The...

Page 435: ...N 5771498 100 REVISION 6 VENUE SERVICE MANUAL P R E L I M I N A R Y 13 Continue with Venue GO EZ Config Setup Wizard EZ Config Setup Wizard Perform the checks listed in Operating System and Applicatio...

Page 436: ...equired Left Side eTower Cover Replacement Procedure No functional checks are required Right Side eTower Cover Replacement Procedure No functional checks are required Mid Thermal Baffle Cover Replacem...

Page 437: ...dure Perform 7 3 3 System Diagnostics Select Full System Diagnostics without JIG Electronic Boards BIB Board Replacement Procedure Perform 7 3 3 System Diagnostics Select Full System Diagnostics witho...

Page 438: ...cs Select Full System Diagnostics without JIG MPB Blower Fan Replacement Procedure Perform 7 3 3 System Diagnostics Select Full System Diagnostics without JIG IPP Module Replacement Procedure Perform...

Page 439: ...nly Installation Procedure Note TheaboveproceduresaredescribedinSoftware Installation Procedure Power on Boot up Power Shut Down Probe Connectors Check 2D Mode B Mode Checks M Mode Checks PW CW Dopple...

Page 440: ...rview of replacement parts for the Venue ultrasound scanner NOTE In the detailed Parts lists illustrations are accompanied by FRU names corresponding Part Numbers and a compatibility matrix This Repl...

Page 441: ...P R E L I M I N A R Y Section 9 2 List of Abbreviations Assy Assembly T BEP Back End Processor T CFE Control Front End CRU Customer replaceable Unit Ctrl Control T FEPS Front End Power Supply LCD Liqu...

Page 442: ...L I M I N A R Y Section 9 3 Main Assemblies and Sub Assemblies Figure 9 1 Venue System View PSU MPB Batteries Covers eTower System Power Distribution Cockpit Monitor Parts Front End Back End Back End...

Page 443: ...Parts Diagram 1 Table 9 5 Mechanical Hardware Parts Diagram 1 Item Part Name Part Number Can Replace Part Can Be Replace d by Part Not Compatib le With Rep l Pro 101 CFE RELEASE ARM S5430823 102 Plast...

Page 444: ...placed by Part Not Compatible With Repl Proc 107 MPB FAN SLEEVE FRU S5755990 108 Base Module FRU S5759996 109 Articular Arm Module FRU S5759998 S5759999 1 110 Articular Arm Module with Locking FRU App...

Page 445: ...9 7 Mechanical Hardware Parts Diagram 3 Item Part Name Part Number Can Replace Part Can Be Replaced by Part Not Compatible With Repl Proc 112 NO LOCK CASTOR S5755602 1 113 DIRECTIONAL and BRAKE LOCK C...

Page 446: ...mpatible With Repl Proc CRU 201 Cover MPB Door S5759973 202 Cover Mid Thermal Baffle FRU S5759974 203 Cover Scanner Door FRU S5759978 204 Cover Right Side eTower FRU S5759982 205 Cover Left Side eTowe...

Page 447: ...Repl Proc CRU 210 Cover Front Base FRU S5759988 211 Cover Left Side Base FRU S5759989 212 Cover Right Side Base FRU S5759990 213 Cover Right Base Bottom FRU S5759991 214 Cover Riser FRU S5759992 215...

Page 448: ...ECTION 5771498 100 REVISION 6 VENUE SERVICE MANUAL Chapter 9 Renewal Parts 9 9 P R E L I M I N A R Y 9 4 3 System Power Distribution Figure 9 7 Power Parts 302 303 304 308 305 311 301 310 307 306 309...

Page 449: ...S5737214 5737214 5737214 302 MODULE MPB MAIN POWER BOARD CRU S5731214 303 AC DC PS CRU S5730963 1 304 MODULE IPP FRU S5730761 305 Cable PSU to MPB DC Docking FRU S5721128 306 Cable ON OFF switch FRU...

Page 450: ...l Parts 9 11 P R E L I M I N A R Y 9 4 4 Cockpit Monitor Parts Figure 9 8 Cockpit Monitor Parts Table 9 11 Cockpit Monitor Parts Item Part Name Part Number Can Replace Part Can Be Replaced by Part Not...

Page 451: ...Cards and Assemblies Item Part Name Part Number Can Replace Part Can Be Replaced by Part Not Compatible With Repl Proc CRU 501 Module T TRX FRU S5764764 502 Module T PSB Probe Selection Board FRU S572...

Page 452: ...10 Back End Table 9 13 Back End Item Part Name Part Number Can Replace Part Can Be Replaced by Part Not Compatible With Repl Proc 601 SSD MODULE FRU S5728757 602 Module BIB BackEnd Interface FRU S5727...

Page 453: ...Part Can Be Replaced by Part Not Compatible With Repl Proc 701 Cable MITX TO SSD PWR FRU S5721127 702 Cable BIB TO BE SPLIT CABLE FRU S5721145 703 Cable BE TO IO USB3 FRU S5715745 704 Cable BE TO IO...

Page 454: ...508 803 Probe L12n RS 5505771 804 Probe 12L RS 5499501 805 Probe E8C RS 5499516 806 Probe 9L RS 5499511 807 Probe C1 5 RS 5499608 808 Probe 6Tc RS KN100106 809 Probe 6S RS 47236956 810 Probe L4 12t RS...

Page 455: ...eplaced by Part Not Compatible With 1001 Power Cable Australia R2415383 7 1002 Power Cable Britain R2415383 4 1003 Power Cable China R2415383 6 1004 Power Cable Denmark 5439668 1005 Power Cable Europe...

Page 456: ...ccessories Figure 9 15 Accessories Table 9 18 Accessories Item Part Name Part Number Can Replace Part Can Be Replaced by Part Not Compatible With Repl Proc 1101 Venue Basic Storage Basket FRU S5767134...

Page 457: ...Optional Peripherals Table 9 19 Optional Peripherals Item Part Name Part Number Can Replace Part Can Be Replaced by Part Not Compatible With Repl Proc 1201 Printer UP D711MD Kit 5449734 1202 Printer...

Page 458: ...Part Number Can Replace Part Can Be Replaced by Part Not Compatible With Repl Proc 1301 ECG Detachable cable AHA Type USA 5146056 1302 ECG detachable cable IEC type EURO and ASIA 5146739 1303 ECG mod...

Page 459: ...re additional tasks and or inspections at a different frequency than listed in this manual 10 1 2 Purpose of Chapter 10 This chapter describes the Care and Maintenance procedures for the Venue ultraso...

Page 460: ...CIRCUIT BOARDS WHILE MAINS POWER TO THE SYSTEM IS ON CAUTION PRACTICE GOOD ESD PREVENTION WEAR AN ANTI STATIC STRAP WHEN HANDLING ELECTRONIC PARTS AND WHEN DISCONNECTING CONNECTING CABLES CAUTION DO N...

Page 461: ...ur GE Service Representative can help you with establishing performing and maintaining records for a Quality Assurance program Contact GE for coverage and or price for service 10 3 3 Maintenance Task...

Page 462: ...ly Per Facility s QA Program Notes Clean Probes or before each use Clean Probe Holders Inspect AC Mains Cable Mobile Ultrasound system Check Weekly Inspect Cables and Connectors Clean Console Clean Co...

Page 463: ...are recommended 2 Bit TX 15 M3 3 Bit TX 20 M4 4 Bit TX 25 M5 5 Bit TX 30 M6 6 Bit TX 45 M10 7 Socket Set must include 7mm socket 8 Side cutter diagonal 5 or 6 inch 9 Flat Blade Driver 3 2 mm 10 Flat...

Page 464: ...ecord all probes and Ultrasound system options 3 Power up Turn the Ultrasound system power ON and verify that all fans and peripherals turn On Watch the displays during power up to verify that no warn...

Page 465: ...affect this mode of operation 2 CF Mode Verify basic CF Mode Color Flow Mode operation Check the basic Ultrasound system controls that affect this mode of operation 3 Doppler Modes Verify basic Doppl...

Page 466: ...te properly 2 B W Printer Verify hardcopy output of the B W video page printer Clean heads and covers if necessary 3 DICOM Verify that DICOM is functioning properly Send an image to a DICOM device 4 E...

Page 467: ...stop the Ultrasound system from moving and release mechanism Where applicable check all wheel locks and wheel swiveling for proper operation 11 Cables and Connectors Check all internal cable harnesses...

Page 468: ...ethanol or methanol alcohol abrasive cleaners or other strong solvents Only use isopropyl alcohol when instructed to do so Do not spray any liquid directly onto the Venue covers Do not allow any liqu...

Page 469: ...touch panel display without affecting the system operation tap Settings and then tap Clean The screen turns black allowing you to use a soft cloth with glass cleaning solution to clean the panel 2 Ta...

Page 470: ...noted Do not drop a probe or strike it against a hard surface as this may damage the probe elements and the acoustic lens or may crack the housing CAUTION TO HELP PROTECT YOURSELF FROM BLOOD BORNE DI...

Page 471: ...es NOTE The most recent up to date information on probes and probe care is available at http www gehealthcare com usen ultrasound products probe_care html Always perform a visual check of the probe pi...

Page 472: ...being visibly clean and properly disinfected The purpose of the regulation is to protect employees in the transportation industry as well as the persons who will receive and or open the package NOTE...

Page 473: ...wing service or repair activities They differ somewhat from the standards that are used for design verification and manufacturing tests e g IEC 60601 1 including national deviations which require a co...

Page 474: ...s holes and similar defects WARNING WARNING Users must ensure that safety inspections are performed whenever damage is suspected and on a regular basis in accordance with local authorities and facilit...

Page 475: ...ms found should be reported to the hospital immediately and the receptacle should not be used NOTE No outlet tester can detect the condition where the Neutral grounded supply conductor and the Groundi...

Page 476: ...to the exposed metal parts of the case The ground wire resistance should be less than 0 2 ohms Reference the procedure in the IEC60601 1 DANGER ELECTRIC SHOCK HAZARD THE PATIENT OR OPERATOR MUST NOT C...

Page 477: ......

Page 478: ...2019 by General Electric Company GE Medical Systems Ultrasound Primary Care Diagnostics LLC 9900 Innovation Drive Wauwatosa WI 53226 USA Tel 1 800 437 1171 Fax 1 414 721 3865 www gehealthcare com...

Reviews: