8-104
Section 8-6 - Mechanical Parts- Replacement Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-6-9
IPP Module Replacement Procedure
8-6-9-1
Tools
Appropriate open-end wrench
Multimeter
Nose pliers
Appropriate short Phillips screwdriver
8-6-9-2
Time Required
10 min
8-6-9-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-6-9-4
IPP Module Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: RS Probe cover, Lower Front eTower, Left side eTower, Right Side
eTower, and MPB Front Metal Door
3.) Remove the MPB Power Door.
4.) Remove battery modules.
5.) Remove the PSU module.
6.) Remove the MPB module.
FRU Part #
Refer to
.
•
Accessories - Replacement Procedures
•
RS Probe Cover Removal Procedure
•
Lower Front eTower Cover Removal Procedure
•
Left Side eTower Cover Removal Procedure
•
Right Side eTower Cover Removal Procedure
•
MPB Front Metal Door Removal Procedure
•
MPB Door Cover Removal Procedure
•
Battery Module Removal Procedure
•
•
Summary of Contents for Venue R1
Page 477: ......