Chapter 8 - Replacement Procedures
8-147
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-7-9
Printer USB Cable Replacement Procedure
8-7-9-1
Tools
Appropriate Phillips screwdriver
8-7-9-2
Time Required
5 min
8-7-9-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-7-9-4
Printer USB Cable Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: RS Probe Cover, Lower Front eTower, Left side eTower, Right Side
eTower, Upper eTower Front Cover.
3.) Remove the MPB Front Metal Door
FRU Part #
Refer to
.
•
Accessories - Replacement Procedures
•
RS Probe Cover Removal Procedure
•
Upper eTower Front Cover Removal Procedure
•
Lower Front eTower Cover Removal Procedure
•
Left Side eTower Cover Removal Procedure
•
Right Side eTower Cover Removal Procedure
•
MPB Front Metal Door Removal Procedure
Figure 8-149 Printer Insert Cover Removed