Site Preparations
2-4
Versana Balance – Basic Service Manual
5808768-100 English Rev.9
2-2-2-2 Environmental specifications for Ultrasound system
The system should be operated, stored, or transported within
the parameters outlined below. Either its operational
environment must be constantly maintained or the unit must be
turned off.
2-2-2-3 Cooling
The cooling requirement for a console Ultrasound system with
monitor and on board peripherals, is up to 3800 BTU/h. This
figure does not include cooling needed for lights, people, or
other equipment in the room.
NOTE:
Each person in the room places an additional 300 BTU/h
demand on the cooling system.
Table 2-2:
System Environmental Requirements
Operational
(Versana Balance with
probe)
Storage
(Versana Balance)
Transport
(Versana Balance)
Temperature
10- 40
°
C
50 - 104
°
F
-5 - 50
°
C
23 - 122
°
F
-5 - 50
°
C
23 - 122
°
F
Humidity
30 - 80% non-condensing
10 - 90% non-condensing
10 - 90% non-condensing
Pressure
700 - 1060hPa
700 - 1060hPa
700 - 1060hPa
CAUTION
Ensure that the probe face temperature does not exceed the
normal operation temperature range.
CAUTION
The Versana Balance system and probe connector is not
waterproof. Do not expose the device to water or any kind of
liquid.
Summary of Contents for Versana Balance
Page 6: ...i 4 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Page 7: ...Versana Balance Basic Service Manual System view i 5 5808768 100 English Rev 9 ...
Page 8: ...i 6 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Page 9: ...Versana Balance Basic Service Manual System view i 7 5808768 100 English Rev 9 ...
Page 10: ...i 8 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Page 11: ...Versana Balance Basic Service Manual System view i 9 5808768 100 English Rev 9 ...
Page 12: ...i 10 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Page 13: ...Versana Balance Basic Service Manual System view i 11 5808768 100 English Rev 9 ...
Page 14: ...i 12 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Page 15: ...Versana Balance Basic Service Manual System view i 13 5808768 100 English Rev 9 ...
Page 16: ...i 14 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Page 17: ...Versana Balance Basic Service Manual System view i 15 5808768 100 English Rev 9 ...
Page 28: ...i 26 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Page 58: ...Introduction 1 30 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Page 390: ...Renewal Parts 9 14 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Page 431: ...GE ...