Section 3-11 - Connectivity Setup - Software v2.x
3-11-8
Verification of a Connection Set Up to EchoPAC PC - Software v2.x
1.) Press
CONFIG
(
F2
) and log on as
adm
4-2-4 "Log On to the System as ‘ADM’" on page 4-12
2.) Select
CONNECTIVITY
(in the lower part of the window).
3.) Select the
VIEWS
tab, located in upper part of window, to display the Current Dataflow Views
.
4.) Select the
DATAFLOW
button to see all available data flows, see
.
5.) Select the
+
beside the dataflow you want to check (here:
Remote Archive - Remote HD
) to see its
contents.
Figure 3-121 Current Dataflow Views screen
Figure 3-122 Pre-configured Data flows in Vivid 7
VIEWS
tab
CONNECTIVITY
tab
DATAFLOW
button
DATAFLOW
button
Select
+