Chapter 4 - Functional Checks
4 - 67
4-3-16
Probe/Connectors Check
NOTE:
Probes can be connected at any time, whether the unit is ON or OFF
CAUTION
Take the following precautions with the probe cables:
- Keep away from the wheels
- Do not bend
- Do not cross cables between probes
Table 4-31 Probe and Connectors Checks
Step
Task
Expected Result(s)
1.
Press
Probe
on the Operator Panel.
A list of the connected probes will pop up on the screen.
2.
If not already selected. use the trackball to select the
desired probe.
An application menu for the desired probe is listed on
the screen.
3.
Trackball to the desired application
Press
Select
to launch the application.
To change application without changing the current
probe, press
Appl.
on the Operator Panel.
The selected application starts.
4.
Verify no missing channels
All channels is functioning.
5.
Verify there's no EMI/RFI or artifacts specific to the
probe.
No EMI/RFI or artifacts.
6.
Test the probe in each active connector slot., see
"Probe Connection" on page 3-21
It will display pictorial data each time
7.
Do a leakage test on the probe, see
"Electrical Safety Tests" on page 10-12
It passes the test.
8.
Repeat this procedure for all available probes.
Summary of Contents for Vivid 7
Page 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Page 2: ......
Page 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Page 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Page 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Page 857: ......