Section 5-13 - Air Flow Control
Section 5-13
Air Flow Control
5-13-1
General Description
The Air Flow Control includes the following components:
•
Dust Filter
•
Dust Filter Frame
•
Air Inlet Channel
•
Fan Assembly
Other components that influence on the airflow is the Front-End Board Rack itself. The air is sucked in
from the rear side of the system through the Filter Cover and the Filter. It passes the Fan Assembly
directly into the upper part of the rear side of the Front-End Board Rack and makes a cooling air current
between the Boards before it blows out against the floor. See
5-13-2
Location in the Unit
The Fan Assy is located between the Board Rack and the Filter, which can be seen on the rear side of
the System.
Figure 5-99 Airflow through the System
FAN ASSY
AIR INLET THROUGH FILTER
AIRFLOW THROUGH BOARD
RACK BEHIND SIDE COVER
AIR OUTLET UNDER THE SYSTEM