Chapter 10 - Care & Maintenance
10 - 7
10-5-2
Functional Checks (See Also Chapter 4)
The functional checks take about 60 minutes to perform. Refer to the system user documentation
whenever necessary.
10-5-2-1
System Checks
NOTE:
* Some software may be considered standard depending upon system model configuration.
Table 10-5 System Functional Checks
Check
Step
Description
2D-Mode
(B-Mode)
Verify basic B-Mode (2D) operation. Check the basic system controls that affect this mode of
operation.
CF-Mode
Verify basic CF-Mode (Color Flow Mode) operation. Check the basic system controls that affect this
mode of operation.
Doppler Modes
Verify basic Doppler operation (PW and CW if available). Check the basic system controls that
affect this mode of operation.
M-Mode
Verify basic M-Mode operation. Check the basic system controls that affect this mode of operation.
*Applicable
Software Options
Verify the basic operation of all optional modes such as Multi-Image, 3D, Contrast, Harmonics,
Cine, Stress Echo,... etc. Check the basic system controls that affect each options operation.
System Diagnostic
Perform the Automatic Tests, see
section Section 7-8 on page 7-63
, to verify that all boards
function according to specifications.
Operator Panel
Test
Perform the Operator Panel Test Procedure, see
keyboard controls are OK. This is performed by the internal PC (Back-End processor) which does
a normal keyboard run through.
Monitor
Verify basic Monitor display functions.
Summary of Contents for Vivid 7
Page 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Page 2: ......
Page 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Page 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Page 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Page 857: ......