3 - 48
Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-6
Set the Scanner’s Network Information - S/W v4.2/v5.2/v6.1/v7.x and later
NOTE:
For software version v4.0/v4.1/v5.0/v5.1/v6.0, see:
3-9-7 "Set the Scanner’s Network
Information - S/W v4.0/v4.1/v5.0/v5.1/v6.0" on page 3-50
.
1.) When in the
Tcpip (
TCP/IP) screen, select
Advanced Settings
to display the
Network
Connections
screen.
2.) Right-click
Local Area Connection
and select
Properties
from the pop-up menu
.
This will display
the
Local Area Connection Properties
screen.
Figure 3-38 Select Advanced Settings
Figure 3-39 Network Connections screen
ADVANCED SETTINGS