3 - 58
Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-14
Set Up Connection to a DICOM Server in a Network
In this case the Vivid 7 is configured to work with DICOM servers in a network environment. Images are
first saved on the local image buffer on the scanner. At the end of the examination the images are sent
to the DICOM server via a DICOM spooler and to the local database, depending on dataflows.
This scenario requires that the scanner is configured to be connected to DICOM servers as described
below.
3-9-14-1
Overview
To connect to the DICOM server, the following information has to be entered in the scanner:
•
The DICOM server IP address.
•
The DICOM server port number.
•
The DICOM server AE title (the server application’s name).
•
Name of Device.
3-9-14-2
DICOM Server IP Address Setting on the Scanner
1.) Press
CONFIG
(
F2
) and log on as
adm
, see
4-2-4 "Log On to the System as ‘ADM’" on page 4-12
2.) Select
CONNECTIVITY
(in the lower part of the window).
3.) Select the
DATAFLOW
tab.
4.) Select the arrow to the right for the Name field to list all dataflows in a pull-down menu, see
5.) Select the dataflow you want to configure from the Dataflow Name pull-down list, see
Figure 3-53 Select Dataflow
ARROW
SELECT DATAFLOW
FROM DATAFLOW NAME
PULL-DOWN MENU.
Summary of Contents for Vivid 7
Page 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Page 2: ......
Page 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Page 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Page 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Page 857: ......