Section 4-2 - General Procedures
Section 4-2
General Procedures
4-2-1
Overview
Some procedures are used more often than other. The intention with this section is to keep the most
used procedures in one place.
NOTICE
SYSTEM REQUIRES ALL COVERS
Operate this unit only when all board covers and frame panels are securely in place. The covers are
required for safe operation, good system performance and cooling purposes.
WARNING
WARNING
ENERGY CONTROL AND POWER LOCKOUT FOR Vivid 7
When servicing parts of the system where there is exposure to voltage greater than 30
Volts: Unplug the system Maintain control of the system power plug There are no test
points to verify isolation, you must wait for at least 20 seconds for capacitors to
discharge Beware that the AC Control Box, Front End Processor and Back End
Processor may be energized even if the power is turned off when the cord is still
plugged into the AC Outlet.
Signed
Date
TAG
&
LOCKOUT