GE H
EALTHCARE
D
IRECTION
FC091194, R
EVISION
11
V
IVID
7 S
ERVICE
M
ANUAL
Chapter 4 - Functional Checks
4 - 39
4-3-7
M Mode Checks
4-3-7-1
Introduction
4-3-7-2
Preparations
Use a phantom when doing these tests.
1.) Connect one of the probes, to the scanner’s left-most probe connector.
-
3-6-5 "Probe Connection" on page 3-21
for info about connecting the probes.
-
Section 9-26 "Probes" on page 9-80
for available probes.
2.) Turn ON the scanner.
-
See see
4-2-2 "Power ON/ Boot UP" on page 4-3
for info about booting the scanner.
The 2D Mode window is displayed (default mode).
3.) Press
M-MODE
to bring up an M-Mode picture on the screen.
Figure 4-29 M-Mode Screen Example