GE H
EALTHCARE
D
IRECTION
FC091194, R
EVISION
11
V
IVID
7 S
ERVICE
M
ANUAL
5 - 6
Section 5-3 - a Front-End Processor (FEP)
Section 5-3
a
Front-End Processor (FEP)
5-3-1
General Description
The electronics in the Front-End Card Cage is also called the Front-End Processor (FEP).
Two versions of the Front-End Card Cage are in use;
•
Version 1 (FEP1) was used from introduction of Vivid 7 until 2nd half of 2003
•
Version 2 (FEP2) was introduced 2nd half of 2003
5-3-2
FEP1
FEP1 has eight
different
types of electronics cards, totally 10 cards. The cards are listed in
NOTICE
Be careful so you don’t insert a card in the wrong position in the card rack. If the power is turned on with
a card placed in the wrong position, the Vivid 7 will be destroyed.
Table 5-2
Front-End Processor Cards used in FEP1
SHORT
NAME
COMPLETE NAME
COMMENT
2X USED PER UNIT
RADIO FREQUENCY & TISSUE BOARD
SPECTRUM DOPPLER PROCESSOR BOARD
SOMETIMES CALLED “IMPORT”
2X USED PER UNIT