Chapter 7 - Diagnostics/Troubleshooting
7 - 69
7-8-3-4
System Test Screen Description
7-8-3-5
Test Loops, System
Figure 7-74 System Test Screen, Upper Part
Table 7-6
Test Loops, System - Test All, System - Break (FEP1)
FIELD
KEY
DESCRIPTION
TEST
LOOPS
-
IN SOME CASES YOU MAY WANT TO DO THE SAME TEST SEVERAL TIMES.
ENTER A NUMBER TO INDICATE HOW MANY TIMES YOU WANT THE TEST TO REPEAT ITSELF.
DEFAULT VALUE = 1..
SYSTEM
TEST ALL
USE
TEST ALL
TO START A SEQUENCE OF ALL THE TESTS AVAILABLE IN THE TEST SCREEN.
THE DIAGNOSTIC TEST IS DONE IN THIS SEQUENCE ON UNITS WITH FEP1;
1.PC2IP
2.IMP (IMAGE PORT BOARD)
3.VME (VME DATA BUS ON THE BACK PLANE)
4.SDP (SPECTRUM DOPPLER PROCESSOR)
5.RFT (RADIO FREQUENCY & TISSUE BOARD)
6.FEC (FRONT-END CONTROLLER BOARD)
7.BF (BEAMFORMER BOARDS)
8.RX (RECEIVER BOARD)
9.TXP (TX POWER)
NOTE:
Relay Board (RLB) and Chain are not tested in this test.
NOTE:
The Transmitter Board (TX) don’t have its own test, but some parts of
the TX board is tested during the FEC test.
NOTE:
If the system must be rebooted, do so. Continue with the testing from
the point at which you were required to stop.
SYSTEM BREAK
IF YOU HAVE STARTED A LOOPED TEST, YOU MAY SELECT
BREAK
TO STOP THE TEST.
THE TEST WILL CONTINUE UNTIL THE CURRENT TEST HAS FINISHED.
System
Test Loops
Mode
Summary of Contents for Vivid 7
Page 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Page 2: ......
Page 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Page 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Page 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Page 857: ......