GE H
EALTHCARE
D
IRECTION
FC091194, R
EVISION
11
V
IVID
7 S
ERVICE
M
ANUAL
8 - 4
Section 8-2 - Side Covers (with Bumpers) Replacement Procedure
Section 8-2
Side Covers (with Bumpers) Replacement Procedure
8-2-1
Manpower
One person, 5 minutes
8-2-2
Tools
Large coin and a large screwdriver
8-2-3
Preparations
Power off / Shutdown the system, see
4-2-3 "Power Shut Down" on page 4-7
.
8-2-4
Side Covers Removal Procedure
1.) Use a large coin to turn the two recessed fasteners at the bottom of the side cover a quarter turn to
the horizontal position.
2.) Take hold under the lower front side of cover and pull gently. At the same time support the rear side
of the cover with the other hand.
NOTE:
The two fasteners hold the left-side cover to the scanner at the bottom of the card cage cover.
The two fasteners hold the right-side cover to the scanner at the bottom of the AC power supply wall.
3.) When the cover is loose in front, pull the rear out and move upwards.
4.) Lift the side cover gently upwards until the three hooks at the top of the side cover are free from the
three retainer slots in the top cover as shown in
.
5.) To Remove Left side cover, repeat this procedure.
NOTE:
shows the right-side cover being removed. Repeat the same steps listed above to remove
the left-side cover.
Figure 8-2 Removal of Side Covers
Coin
Recessed 1/4 turn fasteners
Pull front side loose first
Summary of Contents for Vivid 7
Page 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Page 2: ......
Page 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Page 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Page 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Page 857: ......