8 - 8
Section 8-5 - Air Filter Replacement Procedure
Section 8-5
Air Filter Replacement Procedure
8-5-1
Manpower
One person, 30 minutes
8-5-2
Tools
No special tools are needed
8-5-3
Filter Removal Procedure
1.) Remove Filter Cover, see
8-3-3 "Upper Rear Cover Removal Procedure" on page 8-6
2.) Pull out plastic plugs and remove the filter.
8-5-4
Filter Installation Procedure
Two different Plastic plugs have been used, see
, the one to the left on
early Vivid 7 scanners and the type to the right on newer units (both Vivid 7 and Vivid 7 PRO).
1.) Install a new or a cleaned filter (GE Vingmed Part No: FB307903).
Use plastic plugs as shown to the right in
. (GE Vingmed Part Number
for the plastic plug is: 080X6520.)
2.) Install Filter Cover, see
"Upper Rear Cover Installation Procedure" on page 8-6
.
Figure 8-6 Filter Cover and Filter
Figure 8-7 Plug replacement
Filter
Plastic plugs
Filter Cover
Replace this plug…
with this plug