Section 8-8 - Software Loading Procedure - Software Version v3.x
8-8-11-3
Application SW Installation - Software Version v3.x
1.) Insert the applicable CD labelled
Vivid7 & EchoPAC PC Application SW
BT’03 Software, SW. v3.x" on page 8-39
.
2.) Keep the ON/Standby switch on the operator panel pressed for a few seconds to power down the
scanner. (This is called Forced Shut Down).
3.) Wait until the LCD’s on the Operator Panel is switched OFF and the ON/Standby Switch on the
Operator Panel has turned yellow.
4.) Power ON the System, see
4-2-2 "Power ON/ Boot UP" on page 4-3
5.) A dialogue box, called Start Application, will appear, select
Install SW
.
6.) A dialogue box, called Startloader, will appear, select
OK
to continue.
7.) A dialogue box will appear, also named Startloader stating “New System SW installation will start
now”, select
OK
to continue.
8.) A Command prompt window will appear, hit any key on the alphanumeric keyboard to continue.
9.) A message in the Command prompt window may prompt you that copying application software
takes three minutes. This may take as much as five minutes.
10.)Several dialogue boxes will appear on the screen during the SW load.
11.)Remove the CD from the CD drive when prompted.
12.)Press any key to continue.
13.)The Scanner will now power down and reboot.
WARNING
WARNING
Do not interrupt the SW loading at any time.
Summary of Contents for Vivid 7
Page 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Page 2: ......
Page 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Page 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Page 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Page 857: ......