Chapter 8 - Replacement Procedures
8 - 95
Section 8-15
Front Casters Replacement Procedure
8-15-1
Manpower
1 persons, 30 minutes per Caster
8-15-2
Tools
•
Torx screwdriver T-20
•
Open end wrenches, 10 mm and 134 mm
•
5 mm hexagon key
•
Bevel Edged Board and Wooden Wedge (service parts, see
Table 9-77 "Service Kits, Overview" on
8-15-3
Preparations
1.) Shut down the system, see
4-2-3 "Power Shut Down" on page 4-7
2.) Remove the Front Bumper, see
"Front Bumper Removal Procedure" on page 8-100
3.) Turn Front Casters as shown in
4.) Activate Direction Lock.
5.) Put the Bevel Edged Board on the floor.
Figure 8-91 Bevel Edged Board and Wooden Wedge.
Wooden Wedge
Bevel Edged Board