Section 8-15 - Front Casters Replacement Procedure
8-15-4
Front Casters Removal Procedure
6.) Pull the system backwards up the board incline.
7.) Turn the Rear Caster on the board in the direction as shown in
8.) Activate both Rear Brakes as shown in
WARNING
WARNING
Prior to elevating scanner, verify that the floating Operator Panel is locked in its lowest
position.
WARNING
WARNING
Use extreme caution as long as VIVID 7 is un-stable, not resting on all four Casters.
Figure 8-92 Pull system backwards up the board incline
Figure 8-93 Activate both rear brakes