Chapter 10 - Care & Maintenance
10 - 3
Section 10-3
Maintenance Task Schedule
10-3-1
How often should maintenance tasks be performed?
The Customer Care Schedule (provided on page
) specifies how often your Vivid 7 should be
serviced and outlines items requiring special attention.
NOTE:
It is the customer’s responsibility to ensure the Vivid 7 care & maintenance is performed as
scheduled in order to retain its high level of safety, dependability and performance.
Your GE Service Representative has an in-depth knowledge of your Vivid 7 ultrasound scanning system
and can best provide competent, efficient service. Please contact us for coverage information and/or
price for service.
The service procedures and recommended intervals shown in the Care & Maintenance Task Schedule
assumes that you use your Vivid 7 for an average patient load (10-12 per day) and not use it as a
primary mobile unit which is transported between diagnostic facilities.
NOTE:
If conditions exist which exceed typical usage and patient load, then it is recommended to
increase the periodic maintenance frequencies.
.
Table 10-2 Customer Care Schedule
Service at Indicated Time
Daily
Weekly
Monthly
Annually
Notes
Clean Probes
•*
* or before each use
Clean Probe Holders
•
Clean Air Filter
•
more frequently depending on
your environment
Inspect AC Mains Cable
•
Mobile Unit Check Weekly
Inspect Cables and Connectors
•
Clean Console
•
Clean Monitor and Touch Panel
•
Inspect Wheels, Casters, brakes and
Swivel Locks
•
Mobile Unit Check Daily
Check Operator Panel Movement
•
Mobile Unit Check Daily
Console Leakage Current Checks
•
also after corrective
maintenance or as required by
your facilities QA program
Peripheral Leakage Current Checks
•
also after corrective
maintenance or as required by
your facilities QA program
Surface Probe Leakage Current Checks
•
also after corrective
maintenance or as required by
your facilities QA program
Summary of Contents for Vivid 7
Page 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Page 2: ......
Page 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Page 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Page 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Page 857: ......