GE H
EALTHCARE
D
IRECTION
FQ091013, R
EVISION
1
V
IVID
I
N
AND
V
IVID
Q
N S
ERVICE
M
ANUAL
3-84
Section 3-9 - Configuration
3-9-5
Software Options Configuration
3-9-5-1
Software Option Installation
A Password (Software Option String) enables a software option or a combination of software options.
This password is specific for each Vivid
i
n/ Vivid q N.
3-9-5-2
On-site Configuration
1.) Press
CONFIG
(
F2
) and log on as
Adm
- see
Log On to the System as ADM on page 4-10
.
2.) Select ADMIN (lower part of window),
3.) Select the SYSTEM ADMIN tab.
4.) Select NEW to open the New Key dialog where you type the Software Option Key (Serial String).
5.) Type the Password (Software Option Key).
6.) Press OK to save the new setting.
7.) Close the Config window.
3-9-5-3
Remote Check and Configuration
Contact the OnLine Center for InSite Checkout.
Figure 3-83 System Admin screen
Figure 3-84 Type Software Option Key
CAUTION
Incorrect password entry will result in loss of system options.
If password is incorrect, please contact your local GE Service representative.
System Admin Tab
New
Admin
Field for Software Option Key (Serial String)