GE H
EALTHCARE
D
IRECTION
FQ091013, R
EVISION
1
V
IVID
I
N
AND
V
IVID
Q
N S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-9
8-2-4
Control Panel and Keyboard Replacement Procedure
8-2-4-1
Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the following Control Panel and
Keyboard (and sub-component) replacement procedures.
8-2-4-2
Time Required
10 minutes
8-2-4-3
Preparations
Shut down the Vivid
i
n/ Vivid q N ultrasound unit, as described in
8-2-4-4
Control Panel and Keyboard Removal Procedure
1) Remove the bearing handle, as described in the
Bearing Handle Removal Procedure on page 8 - 3
.
2) Place the Vivid
i
n/Vivid q N portable ultrasound scanner face down and remove the battery as
described in the
Battery Removal Procedure on page 8 - 5
3) Loosen and remove each of the 9 screws that fasten the Control Panel and Keyboard onto the
Bottom Assembly, as indicated in
.
4.) Turn the Vivid
i
n/ Vivid q N over, release the front latches and open the LCD Display cover.
5) Carefully lift the Control Panel and Keyboard upwards from the front, then release it from the rear;
Figure 8-13 Removing the Fastening Screws from the Bottom Assembly
not captive
(do not
This screw
lose!)
Summary of Contents for Vivid in
Page 2: ......
Page 30: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL xxviii...
Page 739: ......