Chapter 8 - Replacement Procedures
8-11
8-2-4-5
Control Panel and Keyboard Installation Procedure
NOTE:
When performing the following installation procedure steps, it may be helpful to review the appropriate
movie clips previously referenced in the
Control Panel and Keyboard Removal Procedure
NOTE:
When installing the Control Panel and Keyboard during other replacement procedures, Step 1 will only
be required if indicated.
1) Stand the replacement Control Panel and Keyboard beside the Vivid
i
n/ Vivid q N system and tilt it
on its side to gain access to the cable connector underneath. Re-connect the keyboard/control
panel cable from the system to the cable connector (refer to
2) Carefully lift the Control Panel and Keyboard into position on the Vivid
i
n/ Vivid q N (refer to
, above). Make sure it is properly seated and take care to align the screw holes with those in
the Bottom Assembly.
3.) Close the Vivid
i
n/ Vivid q N LCD Display cover, making sure both front latches are securely
fastened.
4.) Carefully turn the Vivid
i
n/ Vivid q Nover, placing it face down.
5.) Return and fasten each of the 9 screws that secure the Control Panel and Keyboard onto the
Bottom Assembly (refer to
6.) Re-install the battery as described in the
Battery Installation Procedure on page 8 - 6
7.) Return the Handle as described in the
Bearing Handle Installation Procedure on page 8 - 4
8) Reconnect the AC adapter and all cables to the Vivid
i
n/Vivid q N portable ultrasound scanner.
9.) Proceed to perform the following functionality tests:
-
SafeLock Cart - Grounding Continuity on page 10 - 23
- only if the Vivid
i
n/ Vivid q N system
is currently mounted on the SafeLock cart
-
SafeLock Cart - Chassis Current Leakage Test on page 10 - 25
-
Isolated Patient Lead (Source) Leakage – Lead-to-Ground on page 10 - 27
-
Isolated Patient Lead (Source) Leakage – Lead-to-Lead on page 10 - 28
-
Isolated Patient Lead (Sink) Leakage - Isolation Test on page 10 - 28
-
Alphanumeric Keyboard Test on page 4 - 19
-
External Keyboard Test on page 4 - 20
-
-