Section 8-2 - External Component Replacement Procedures
6.) Check that when closing the LCD Display cover it aligns properly with the Bottom Assembly. If
necessary, make further adjustments.
7.) Refit the LCD Display Frame, as described in the
LCD Display Frame Installation Procedure on
8.) Return and fasten the 2 screws (previously removed) to each of the LCD hinges, making sure each
hinge is securely fastened to the Base Assembly - refer to
Do not over-tighten the screws
.
9.) Reconnect the LCD flex cable to the BEP (refer to
).
10.)Refit the Control Panel and Keyboard, as described in the
Control Panel and Keyboard Installation
.
11.)Close and open the lid a few times to ensure proper closure of the LCD Display cover.
12) Reconnect the AC adapter and all cables to the Vivid
i
n/Vivid q N portable ultrasound scanner.
13.)Proceed to perform the following functionality tests:
-
SafeLock Cart - Grounding Continuity on page 10 - 23
- only if the Vivid
i
n/ Vivid q N system
is currently mounted on the SafeLock cart
-
SafeLock Cart - Chassis Current Leakage Test on page 10 - 25
Figure 8-29 LCD Display Hinges Secured to Base Assembly - Location of Screws
Here