Chapter 8 - Replacement Procedures
8-27
8-2-8
LCD Display Frame Replacement Procedure
8-2-8-1
Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the bottom assembly replacement
procedures.
8-2-8-2
Time Required
10 minutes
8-2-8-3
Preparations
Shut down the Vivid
i
n/ Vivid q N ultrasound unit, as described in
8-2-8-4
LCD Display Frame Removal Procedure
1) With the LCD Display in the open (upright) position, use a flat screwdriver to remove the screw cap
from each of the four screws that fasten the LCD display frame in position, as shown in
.
Figure 8-30 Removing the Screw Caps from the LCD Display Frame
2) Loosen and remove the four screws (
).
3) With the LCD Display in the fully open position, carefully release the LCD display frame from the
Figure 8-31 Removing the Securing Screws