Chapter 3 - System Setup
3-5
Section 3-3
Receiving the
Vivid
i
n and Vivid q N
3-3-1
Examin All Packages
Examin all packages closely at time of delivery, as described in the procedure below.
Table 3-4
Examin All Packages
STEP
TASK
ILLUSTRATIONS
1
Is dammage apparent?
•
If
yes
; continue with the instructions in subsection
3-3-2 - Damage in Transportation
•
If no; continue with step 2.
2
Is the Shock Indicator red colored inside the middle of the
indicator?
•
If
yes
: The Shock Indicator has been trigged.
Make a remark on the Post Delivery Checklist about the
trigged indicator before you continue with step 2.
•
If
no
: Continue with step 2.
3
Is the Tilt Indicator red colored inside the middle of the
indicator?
•
If
yes
: The Tilt Indicator has been trigged.
Make a remark on the Post Delivery Checklist about the
trigged indicator and then follow the rest of the
instructions in subsection
•
If
no
: Continue with the instructions in
Receiving the Vivid i n and Vivid q N
WARNING
WARNING
For Vivid i n and Vivid q N portable ultrasound scanners supplied from the factory with
a dummy battery, when unpacking the unit, do not discard the dummy battery.
When charging the Vivid i n or Vivid q N lithium ion battery on an external charger, since
the battery in fact forms one of the four legs of the scanner, removing it will leave the
unit standing unbalanced. For this reason, the dummy battery should be kept, so it may
be inserted in position to provide stability to the scanner while the lithium ion battery
is being charged.
In addition, use of the dummy battery is recommended during transportation or
long-term storage of the Vivid i n and Vivid q N portable ultrasound scanner.
RED COLOR
RED COLOR