GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
3-48
Section 3-7 - Completing the Hardware Installation
3-7-2-8
Connecting the VCR
There are two types of VCR - PAL or NTSC approved by GE Medical Systems.The VCR connection in
the different scenarios is explained below.
•
Direct Connection to the Vivid q N System
A Mitsubishi VCR is used in this configuration - use the type that is suited to your region. The video
signal is received through a VGA-to-video converter.
1) Connect one end of the VGA cable to the VCR OUTPUT located on the Vivid q N rear connector
panel (see note below).
NOTE:
When performing
, it may be necessary to use the VGA adaptor (P/N 2420728 - see
) to connect the end of the VGA cable to the Vivid q N rear connector panel.
Since some VGA plugs have thicker housing, they do not fit properly into the
Vivid q N VGA connector socket on the rear panel (because the plug is not centrally aligned
within the frame). In this event, use of the VGA adaptor will eliminate the possibility of causing
damage to the BEP by the use of unnecessary force.
2) Connect the other end of the VGA cable to the VGA INPUT on the video converter - see
.
3.) Connect one end of the S-Video cable to the S-Video OUTPUT on the video converter.
Figure 3-43 Video Casette Recorder
Figure 3-44 VGA Adaptor
Figure 3-45 Video Converter