GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
3-68
Section 3-8 - Mounting the Vivid q N on the SafeLock Cart (optional)
3-8-2
Adjusting the SafeLock Cart Locking Mechanism
NOTE:
The alignment procedure below is recommended in the event that when mounting the Vivid q N
portable ultrasound scanner on the SafeLock Cart, the unit does not lock firmly in position when
inserted into the safelocks. Alternatively, if when removing the unit from the Cart, it is difficult to
disengage the safelocks.
1.) Working from the left side of the SafeLock Cart, remove the DVD from the DVD/CD-RW device.
Working from underneath the main shelf, loosen and remove the four screws (two each side) from
the DVD/CD-RW device bracket, then remove the bracket and the device - see
.
2.) Working from the underneath the main shelf, use a Phillips screwdriver to loosen the two screws
(half turn counter-clockwise) holding the middle locking tooth in position on both the left and right
sides of the main shelf, as shown below.
3.) Check that the Vivid q N portable ultrasound scanner is locked correctly onto the two rear locking
teeth (if not, push the scanner fully back into position until an audible click is heard).
4.) Carefully tighten the four screws previously loosened in
.
5.) Reconfirm that the scanner is firmly locked into position and there is no excess play (movement)
between the scanner and the Cart before returning the DVD/CD-RW device and bracket.
Figure 3-66 Figure 3-85--DVD/CD Device and Bracket
Figure 3-67 Figure 3-86--Adjusting the SafeLock Cart Locking Mechanism
Note
: only two screws
visible in picture
Screw
Screw
Middle Locking Teeth Viewed from Top of Shelf
Screws Viewed from Beneath Shelf