Chapter 5 - Components and Function (Theory)
5-33
Recommended storage conditions of battery pack:
Short term (less than one month):
0° C (32 °F)
to 50 °C (122 °F)
Long term (more than three months): 10° C (50 °F) to 35 °C (95 °F).
CAUTION
The battery has a safety device.
• Do
not
disassemble or alter it. Charge the batteries only when the ambient temperature is between
0° and 65° C (32° and 149 °F) and discharge the batteries between -10° and 55° C (14° and 131 °F).
• Do
not
short-circuit the battery by directly connecting the battery terminals with metal objects.
• Do
not
heat the battery or incinerate.
• Do
not
expose the battery to temperature over 60° C (140° F). Keep it away from fire and other heat
sources.
• Do
not
charge the battery near a heat source, e.g. fire or heaters.
• Do
not
leave the battery in direct sunlight.
• Do
not
pierce the battery with a sharp object, hit it, or step on it.
• Do
not
use a damaged battery. Do not solder a battery.
• Do
not
connect the battery to an electrical outlet.
CAUTION
If the Vivid q N portable ultrasound system is not being used for periods of one month or longer, the
battery should be removed during lengthy non-use period.
CAUTION
To avoid the battery bursting, igniting, or fumes from the battery causing equipment damage,
observe the following precautions:
• Do
not
immerse the battery in water or allow it to get wet.
• Do
not
put the battery into a microwave oven or pressurized container. If the battery leaks or emits
an odor, remove it from all possible flammable sources.
• If the battery emits an odor or heat, is deformed or discolored, or in a way appears abnormal during
use, recharging or storage, immediately remove it and stop using it.
• If you have any questions about the battery, consult GE or your local representative.