Chapter 6 - Service Adjustments
6-1
Chapter 6
Service Adjustments
Section 6-1
Overview
6-1-1
Purpose of Chapter 6
This section provides references to the procedures required for testing and adjusting the Vivid q N portable
ultrasound scanner. These tests are optional, and may be used to check the system for errors.
NOTE:
There are no voltage adjustments required.
6-1-2
LCD Display Adjustments and Calibration
Overall low image quality or “shades” visible on the Vivid q N LCD Display may be the result of improper
monitor calibration. Therefore, adjustments may be made to the
Contrast, Brightness
,
Watermark
and/
or
Blue Tint
, as required.
Important
: LCD Display adjustments should be made in the correct order, as stated in the following
procedures.
NOTE:
Incorrect calibration of the Vivid q N monitor may also result in less than optimal image quality
when reviewing images on a DICOM or EchoPAC station.
Access the Calibration Screen (by pressing <
ALT
>+<
Q
>), then follow the relevant procedure
instructions described below.
.
NOTE:
The Hot keys used to exit the Calibration Screen are: <
ALT
>+<
Q
>.
Figure 6-1 LCD Display Calibration Screen