Chapter 8 - Replacement Procedures
8-17
8-2-4-5-9
ECG/Respiratory Module Assembly Removal Procedure
NOTE:
For Vivid q N systems that previously did not have the Respiratory Option, it is necessary
remove the ECG Board from the ECG Module Assembly and replace it with an
ECG/Respiratory Board. The preparation and removal procedures below describe the required
steps.
Preparation
1) Access the
D:\
directory and delete the file named traces.res, as shown below:
NOTE:
In the event that the traces.res file is not deleted prior to installing the new ECG/Respiratory
board, the following error message will later be displayed:
2.) Shut down the Vivid q N ultrasound unit, as described in
Removal Procedure
1) Lift the Control Panel and Keyboard up to gain access to the cable connector underneath and
disconnect the keyboard/control panel cable from the system - refer to
2.) Loosen and remove the four screws (two long and two short) on either side of the bracket that fasten
the ECG Module assembly to the base of the Control Panel, as shown in
.
Figure 8-21 ECG Module Assembly Removal
D:\Marlin\Test\resources\Nemo\acquis\HWConfig\
traces.res