Chapter 8 - Replacement Procedures
8-19
8.) Press
Config
"Log On to the System as ADM" on page 4-10
9.) Select
About
, then click on the
HwVersion
tab.
10.)Verify that the correct ECG/Respiratory module Part # is displayed, as shown in
11.)Proceed to perform the following functionality tests:
-
"Isolated Patient Lead (Source) Leakage – Lead-to-Ground" on page 10-26
-
"Isolated Patient Lead (Source) Leakage – Lead-to-Lead" on page 10-27
-
"Isolated Patient Lead (Sink) Leakage - Isolation Test" on page 10-27
-
"SafeLock Cart - Grounding Continuity" on page 10-22
- only if the
Vivid q N system is currently mounted on the SafeLock cart
-
"SafeLock Cart - Chassis Current Leakage Test" on page 10-24
Figure 8-23 Hardware Version Verification