GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
8-70
Section 8-4 - SafeLock Cart Components Replacement
8-4-5
ECG Cable Protector Replacement Procedure
8-4-5-1
Tools
Appropriate Phillips screwdriver.
8-4-5-2
ECG Cable Protector Removal Procedure
1.) While facing the SafeLock Cart (i.e., working from the front of the Cart,) loosen and remove the two
screws that secure the ECG Cable Protector to the Top Shelf Assembly (refer to
). See
2.) Remove the ECG Cable Protector from the Top Shelf Assembly.
8-4-5-3
ECG Cable Protector Installation Procedure
If this is a first-time installation of the ECG Cable Protector, it is necessary to remove the two plastic
guide holders from the Top Shelf Assembly, as shown in
. While facing the SafeLock Cart
(i.e., working from the front of the Cart,) these are located on the right side of the SafeLock Cart.
1.) Thoroughly clean the surface of the Top Shelf Assembly, in particular the recessed area from where
the ECG Cable Protector (or plastic guide holders) were removed.
2.) Position the ECG Cable Protector on the Top Shelf Assembly surface, ensuring that the recessed
groove on the protector aligns with the groove on the SafeLock Cart, as shown in
Figure 8-82 Removing Plastic Guide Holders from Top Shelf Assembly
Plastic guide holders
on right hand side