Chapter 8 - Replacement Procedures
8-89
8-4-14
Top Shelf Replacement Procedure
8-4-14-1
Tools
Appropriate Phillips screwdrivers, Allen key and open wrench.
8-4-14-2
Preparations
Shut down the Vivid q N ultrasound unit, as described in
8-4-14-3
Top Shelf Removal Procedure
1) Disconnect the AC adapter and all cables from the Vivid q N portable ultrasound scanner.
2.) Disconnect and remove all peripherals from the SafeLock Cart.
3.) Remove the Upper Cover (Rear) as described in the
"Upper Cover (Rear) Removal Procedure" on
4.) Remove the Lower Cover (Rear) as described in the
"Upper Cover (Rear) Removal Procedure" on
5) Remove the USB board as described in the
"USB Board Removal Procedure" on page 8-82
6) Remove the gas spring as described in
"Gas Spring Replacement Procedure" on page 8-84
7) Disconnect the two cables (P1 cable and peripheral power cable) from the top of the AC Input
Assembly as described in
"AC Distribution Assembly Removal Procedure" on page 8-74
(refer to
and
).
8) Remove Main Harness, Peripheral power cable and disconnect the LAN cable.
9) Remove the six lower screws that retain the top Shelf - see
.
10) Remove the two rear screws that retain the top Shelf - see
.
11) Carefully remove the top shelf assembly.
Figure 8-104 Top Shelf - Lower Retaining Screws
Lower screws retaining the
Top Shelf - 6 shown