Section 10-6 - Electrical Safety Tests
10-6-8-1
Data Sheet for ECG Current Leakage
The test passes when all readings measure less than the values shown in the tables below. Record all
data.
Table 10-17 Maximum Allowance Limit for ECG Lead Leakage Test to Ground
AC Power
Source
Maximum
Allowance Limit
GROUND
OPEN
GROUND
CLOSED
Patient Lead-to-Ground Current Leakage Test
and
Patient Lead-to-Lead Current Leakage Test
115V
10uA
10uA
220/240V
500uA
10uA
Table 10-18 Maximum Allowance Limit for ECG Lead Isolation Sink Test
AC Power
Source
Maximum
Allowance Limit
Patient Lead Isolation Current Test
115V
20uA
220/240V
5mA