GE H
EALTHCARE
DRAFT
V
OLUSON
E8 / V
OLUSON
E6
D
IRECTION
KTD102576, R
EVISION
7
DRAFT (A
UGUST
23, 2012)
S
ERVICE
M
ANUAL
Chapter 4 - Functional Checks
4-43
5.) Choose archiving span time from the pop-up menu.
6.) If desired, check mark “Remove Local Images after Backup”.
7.) Click the ADVANCED button if it is desired to adapt archive data.
8.) Select the NEXT button.
9.) Choose the Destination.
10.)Enter the description of the image backup.
11.)Select the NEXT button.
12.)To start the backup process click YES.
NOTICE
!! NOTICE:
If for example “All images older than”
1 Day
is chosen (see
), images of the
current day will not be archived! However, if you click the [Advanced] button you can put this right.
Figure 4-37 Image Archive Save - choose destination
CAUTION
!! CAUTION:
Voluson E8 / Voluson E6 presets, configurations and image settings will not be backed up! To
Backup the Full System Configuration
choose destination
enter backup
click [Next] button
description
Summary of Contents for Voluson E8