Section 5-15 - Service Page
Section 5-15
Service Page
5-15-1
Introduction
The Service Page contains specific software/hardware test modules, system setup, update, etc. for
Voluson systems only.
5-15-2
Access / Security
The service page has different access and security user levels.
Each user is only granted access to the tools that are authorized for their use.
5-15-3
Service Login
1.) Press the
UTILITIES
key on the control panel.
2.) In the “Utilities” menu touch SYSTEM SETUP to invoke the setup desktop on the screen.
3.) On the left side of the screen select ADMINISTRATION and then click the
SERVICE
tab.
4.) Enter the password
SHE
and click the ACCEPT button to display the Service Tools window.
5-15-3-1
Auto Tester
Autotester is a log function of customer activities. It records all user actions (scanning, Touch Panel
entries, performing Calculations, review of Patient Reports, etc.…). It is possible to safe (record) as file
on HDD. But also export to DVD/CD can be done to allow replay of the records on other systems.
Section 7-6 "How to use the Auto Tester program" on page 7-16
.
Figure 5-49 Service Tools window
NOTICE
!! NOTICE:
For intermittent problems this file can be requested from the Service Engineer or customer.
It is possible to burn this file on DVD/CD+R/RW.
Activate Service
Export System
Auto Tester
Monitor Test
Delete all patients
Update
TCP/IP Buffersize
Common
Keyboard Layout
Service Desktop
Exit
Data
Request for
Service
Load Application
Settings