Section 5-16 - Boot Screen Functions
5-16-1-3
Rollback
This function offers the possibility to simply restore the previously used system configuration (rollback),
which was stored on R:\ during “FMI from DVD”.
1.) Turn system OFF and then back ON.
2.) As soon as the “Boot Screen” appears (see:
), press the
PG DN
[
↓
Arrow down] key on the keyboard until the
Rollback
item is highlighted, then press
ENTER
.
3.) When the following WARNING message appears, press the
← [
Arrow left] button to highlight OK
and then press
ENTER
.
After performing the rollback, the system reboots.
The Ultrasound Application is started and finally the 2D screen is displayed on the monitor.
5-16-1-4
Memtest
Starts the PC Memory Test.
Operation see:
Section 5-16-2 "Memory Check in LINUX" on page 5-77
Figure 5-53 Warning message
select [OK]