Chapter 8 - Replacement Procedures
8-55
8-19-11
Replacement of the Probe Holder for Endocavity probes
8-19-11-1
Manpower
One person, 5 minutes
8-19-11-2
Tools
Phillips screwdriver 1
8-19-11-3
Preparations
none
8-19-11-4
Probe Holder- Removal Procedure
1.) Screw off the 2 screws and carefully remove the Probe Holder.
8-19-11-5
Probe Holder- Installation Procedure
1.) Place the new Probe Holder and fasten it with the screws as shown in
Figure 8-75 on page 8-55
.
Figure 8-75 screw off the screws