Vscan Air Maintenance
6-6
Vscan Air
–
User Manual
Direction GP092020-1EN
Rev 18
Cleaning and disinfection
Reprocessing recommendation (Frequency)
After Each Use
1. Inspect the Vscan Air CL (See ‘Inspecting the Vscan Air’ on
2. Clean the Vscan Air CL.
3. If required, clean the display device.
NOTE:
The display device should be cleaned and/or disinfected
according to the device manufacturer's recommendations.
Example: https://support.apple.com/en-us/HT204172.
Please be aware that medical grade display devices often
support a wider range of cleaners and disinfectants
compared to display devices for the consumer market.
4. If required, disinfect the Vscan Air CL.
Ensure that the Vscan Air CL is properly cleaned and
disinfected after each use and before storage in the
protective case.
WARNING
If any defects or damages are found on the Vscan Air CL,
do not use it. Contact GEHC service.
WARNING
Risk of Infection. ALWAYS clean and disinfect the probe
between patients to the level appropriate for the type of
examination.