Preparing Vscan Air CL for Use
3-12
Vscan Air
–
User Manual
Direction GP092020-1EN
Rev 18
Wireless connection quality indicator (continued)
The ‘Wireless connection quality indicator’ provides a visual
indication of the connection quality between the probe and the
app during scanning. An unstable connection may result in loss
of image quality or slow image update during real time imaging.
Probe temperature indicator and thermal management
The ‘Probe temperature indicator’ tracks and displays changes
in the operating temperature of the probe during scanning.
Factors affecting probe temperature are:
•
Transducer: Curved array transducer gets warm more
quickly than linear array due to higher power consumption.
•
Preset: Certain presets like Abdominal and cardiac have
higher power requirements depending on the image settings
causing the probe to warm up faster than other presets.
•
Mode: Operating in Color flow mode warms up probe faster
than Black and white mode.
•
Length of scan: Duration of continuous scanning.
•
Ambient temperature: Higher ambient temperatures can
cause the probe to warm faster.
1. Stable
connection
2. Unstable
connection
1. Thermal
indicator
Summary of Contents for Vscan Air
Page 5: ...Vscan Air User Manual i 3 Direction GP092020 1EN Rev 18 This page intentionally left blank...
Page 14: ...i 12 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 324: ...Using Vscan Air CL 5 112 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 370: ...Vscan Air Maintenance 6 46 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 382: ...Privacy and Security 7 12 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 406: ...Index 2 Vscan Air User Manual Direction GP092020 1EN Rev 18...
Page 407: ......