6 Maintenance and service
Cleaning
Keep the WAVE Cabinet dry and clean. Use appropriate cleaning detergent on the
stainless steel surfaces of the WAVE Cabinet. Take great care not to spill anything on
the movable parts of the shelves or on any electrical equipment.
Changing fuse
WARNING
Disconnect power. Always disconnect power from the instrument
before performing any installation or maintenance task.
Action
Step
Turn off the instrument.
1
Disconnect the power cord.
2
Lift out the fuse holder, located between the power cord connection and
the switch.
3
Check/change the two fuses in the fuse holder. Use only T10AL250V certified
fuses.
4
Insert the fuse holder.
5
Connect the power cord. Always use a grounded power outlet.
6
Turn on the instrument.
7
20
WAVE Cabinet 20/50 User Manual 28-9651-86 AA
6 Maintenance and service
Summary of Contents for WAVE Cabinet 20