4.4.5
Temperature screens
Temperature Detail screen
The Temperature Detail screen is opened by tapping TEMP. on the navigation toolbar.
The temperature probe contains two sensors. The data from both sensors is displayed
on the Temperature Detail screen.
The following illustration shows the Temperature Detail screen.
1
2
3
4
5
6
7
8
Function
Description
Part
Temperature setpoint (SP), the designated
temperature of the disposable bag contents.
Text box for input.
TC_163 Temperature SP
1
Current temperature (process variable, PV)
displays for temperature probe sensors 1 and
2. The currently selected sensor is shown in
black.
TE_161 Temperature PV
TE_162 Temperature PV
2
Temperature PID controller output (controlled
variable, CV) display, shown in % of config-
ured TCU remote setpoint temperature range.
Note:
In Auto mode this value is calculated by the
PLC. In Manual mode, TC_163 Manual CV is
used as TC_163 Temperature CV.
TC_163 Temperature CV
3
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD