Action
Step
Use the navigation keys to make selections in the menu:
6
a
Select Calibrate and press ENTER (2).
b
Select Sensor 2 and press ENTER.
c
Select Cell K.
d
Use the navigation keys to enter the cell constant value that you
found in the accompanying manual.
e
Press ENTER.
f
Press EXIT (3).
Result: The main screen is displayed. The conductivity probe is ready to be
used.
Rinse the probe with de-ionized water.
7
Disconnect the probe cable from the probe.
8
This procedure is sufficient for most applications. If additional procedures are needed
(changing an offset or resetting the system), refer to the manufacturer's documentation.
164
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
6 Preparation
6.4 Prepare and install the probes
6.4.2 Prepare the conductivity probe