10.7.1
TCU requirements
TCU requirements - 50 L, 100 L,
200 L mixer
Value
Parameter
Property
5 kW
Cooling capacity
General requirements
5 kW
Heating capacity
2.8 bar (40 psi g)
Fluid outlet pressure
Process requirements
40 L/min
Fluid outlet flow
-10°C
Fluid outlet minimum temperature
65°C
Fluid outlet maximum temperature
1.5
″
Connector size
4 to 20 mA
Setpoint
Remote control interface
(for XDUO mixers only)
Required
On/Off emergency stop
TCU requirements - 500 L, 1000 L
mixer
Value
Parameter
Property
9 kW
Cooling capacity, 500 L mixer
General requirements
15 kW
Cooling capacity, 1000 L mixer
9 kW
Heating capacity, 500 L and 1000 L mixer
2.8 bar (40 psi g)
Fluid outlet pressure
Process requirements
60 L/min
Fluid outlet flow
-10°C
Fluid outlet minimum temperature
65°C
Fluid outlet maximum temperature
1.5
″
Connector size
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD