Action
Step
Set the mixer in Auto mode: drag the toggle button
to Auto.
4
Result: The pH of the disposable bag contents is controlled by acid and base
pumps to the given setpoint.
Work in Manual mode
In Manual mode, the manually defined CV (pHC_125 Manual CV) range 0% to 50%
controls the CV of the acid pump (SC_201), and the CV range 50% to 100% controls the
base pump (SC_202). See the table below.
SC_202 Base CV (%)
SC_201 Acid CV (%)
pHC_125 Manual CV (%)
0
100
0
0
80
10
0
60
20
0
40
30
0
20
40
0
0
50
20
0
60
40
0
70
60
0
80
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
217
7 Operation
7.3 Operate Xcellerex XDUO mixer
7.3.4 Control the pH