Note:
Agitation, Xcellerex XDM mixer
Corrective action
Possible causes
Problem
Acknowledge the agitator fault
and attempt to restart the mo-
tor. If the error persists, contact
GE for service.
The motor is overloaded.
Variable frequency drive (VFD) is
not functioning as expected.
Connect the power cable.
Power cable is not connected.
Contact your GE representative
to replace the drive.
The drive is broken.
Refer to the VFD manufacturer's
manual for error code clarifica-
tion. If the problem cannot be
solved on site, contact your GE
representative for help.
Software or mechanical fault in
the VFD.
An error code is shown on the
VFD display.
Turn the power switch to I ON
position.
The power switch on the I/O
Cabinet is in O OFF position.
The agitator motor cannot be
started.
Connect the power cable.
Power cable is not connected.
Contact your GE representative
to replace the drive.
The drive is broken.