Agitation, Xcellerex XDUO mixer
Corrective action
Possible causes
Problem
Check the weight of the dispos-
able bag. See Section 2.4
Xcellerex XDUO mixer interlocks,
on page 43 for more informa-
tion.
Min Weight interlock is active.
The agitator does not start.
Tap RESET on the local control
software home screen and start
the agitator when the system
has been restarted after emer-
gency shutdown.
Note:
The agitator cannot be started
while the system is in the state
E-STOP.
The system is in the state of E-
STOP.
Enable the agitator. See
FlexFactory Automation User
Manual for more information.
The agitator is not enabled.
The agitator does not start on
FlexFactory controlled mixer.
Tap RESET on the local control
software screen and start the
agitator when the system has
been restarted after emergency
shutdown.
Note:
The agitator cannot be started
while the system is in the state
E-STOP.
The system is in the state of E-
STOP.
1
Set the system to Auto
mode.
2
Specify the setpoint.
Note:
When the system is in Manual
mode the setpoint is ignored.
The system is in Manual mode
and cannot accept the setpoint.