Corrective action
Possible causes
Problem
Contact your GE representative
to schedule agitator alignment
check and maintenance.
The agitator is not properly
aligned.
The agitator assembly does not
fit properly in place.
The agitator drive head rubs
against the agitator base plate
support and a grinding sound is
emitted. Plastic filings might be
generated.
Burst disc
Corrective action
Possible causes
Problem
Replace the burst disc. See Re-
place burst disc, on page 259 for
instructions.
Burst disc has ruptured.
Liquid is leaking from the vessel
jacket.
Local control panel
The information presented in this section is relevant for XDUO mixer only.
Corrective action
Possible causes
Problem
Connect the power cable.
Power cable is not connected.
The local control panel does not
function.
Contact your GE representative
for replacement cable.
The main power cable is dam-
aged.
Contact your GE representative
for possible replacement.
The I/O Cabinet is broken.
The home screen does not show
up on the local control panel, al-
though the system is plugged in.
Contact your GE representative
for possible replacement.
The local control panel screen is
broken.
A button on local control panel
screen is not working.
Contact your GE representative
for assistance with configuration
adjustment.
The mixer configuration is not
correctly set up.
A specific screen is not accessi-
ble from the local control panel.