Pumps
The information presented in this section is relevant for XDUO mixer only.
Corrective action
Possible causes
Problem
Tighten the tubing clamps in the pump
head.
The tubing clamps in the pump head
are not properly set for the installed
tubing.
The tubing is
moving in the
pump head.
Adjust the biting elements using the
thumbscrews on the lower right and left
sides of the pump.
The tubing is not fastened in the pump
head (pump model 313).
Contact your GE representative for re-
placement.
•
Loose wiring
•
Failed motor
•
Broken gearbox
The pump drive
head is not turn-
ing.
Move the section of worn out tubing to
the left or right.
All tubing used in the pumps will wear
out with use over time.
The tubing is worn
out.
Use only tubing rated for use in peri-
staltic pump head.
Verify that the cause of the E-STOP
condition is eliminated, and clear the E-
STOP. Try to use the pumps again.
The system is in E-STOP state.
The pump is not
turning.
Clear the interlock condition.
An interlock is active. See Section 2.4
Xcellerex XDUO mixer interlocks, on
page 43 for more information.
Temperature control
The information presented in this section is relevant for jacketed XDM and XDUO mixers.
Troubleshooting for most common TCU models is described below. If your TCU model is not described,
contact your GE representative for help with troubleshooting.
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD